Evaluation of Volumetric Losses During Radiation Therapy Using Image Guidance of Electronic Portal Imaging Device

Loading...
Thumbnail Image

Date

2010

Advisors

Chang, Zheng

Journal Title

Journal ISSN

Volume Title

Repository Usage Stats

415
views
259
downloads

Abstract

Purpose: Changes in patient volume, due to tumor shrinkage, dehydration, dysphagia and atrophy, could present issues in the accuracy of dosimetry throughout the course of treatment. The aim of this work is to study the dosimetric impacts of the volumetric changes during IMRT and to investigate the feasibilities of electronic portal imaging device (EPID) in predicting the impacts. Materials and Methods: An anthropomorphic head and neck phantom was used to represent two scenarios: symmetric and asymmetric volume loss. The phantom was simulated and planned according to the head and neck protocols used in our clinic. Dose volume histograms (DVH) were generated for each set up scenario and were used to calculate the integral dose expected at the coincident volume of the phantom. During treatment delivery, the EPID captured exit fluence of each beam at each level of bolus thickness. These images were quantitatively analyzed using gamma analysis with criteria of 3% and 3mm dose difference and distance-to-agreement respectively. Results: Comparing maximum to minimum volume in the symmetric situation with DVH generated in Eclipse show substantial fluctuations in dose. When comparing five layers of bolus material to zero layers of bolus material, the changes were most significant. The asymmetric volume change predicted dose fluctuations that were less significant than the symmetric phantom. As for gamma analysis, a quantitative evaluation of the integrated dose fluence, captured by the EPID, showed extreme variability in the images with five layers of bolus when compared to images with no bolus. Less significant variation was shown in layers of closer thicknesses, as expected. Conclusions: The phantom study indicates that volume loss could contribute to clinically considerable changes in the dose delivered to target and organs at risk. The proposed technique using EPID could provide valuable information about the variation of dose due to volumetric changes and might be potentially useful.

Description

Provenance

Citation

Citation

Senick, Scott Michael (2010). Evaluation of Volumetric Losses During Radiation Therapy Using Image Guidance of Electronic Portal Imaging Device. Master's thesis, Duke University. Retrieved from https://hdl.handle.net/10161/3062.

Collections


Dukes student scholarship is made available to the public using a Creative Commons Attribution / Non-commercial / No derivative (CC-BY-NC-ND) license.