Chang, ZBowsher, JCai, JYoo, SWang, ZAdamson, JRen, LYin, FF2019-10-042019-10-042012-07-051526-99141526-9914https://hdl.handle.net/10161/19403American Association of Physicists in Medicine (AAPM) task group (TG) 142 has recently published a report to update recommendations of the AAPM TG 40 report and add new recommendations concerning medical accelerators in the era of image-guided radiation therapy (IGRT). The recommendations of AAPM TG 142 on IGRT are timely. In our institute, we established a comprehensive imaging QA program on a medical accelerator based on AAPM TG 142 and implemented it successfully. In this paper, we share our one-year experience and performance evaluation of an OBI capable linear accelerator, Novalis Tx, per TG 142 guidelines.HumansImage EnhancementRadiotherapy Planning, Computer-AssistedCalibrationParticle AcceleratorsQuality ControlRadiotherapy, Image-GuidedImaging system QA of a medical accelerator, Novalis Tx, for IGRT per TG 142: our 1 year experience.Journal article2019-10-04