Item Open Access Accuracy and efficiency of image-guided radiation therapy (IGRT) for preoperative partial breast radiosurgery.(Journal of radiosurgery and SBRT, 2020-01) Yoo, Sua; O'Daniel, Jennifer; Blitzblau, Rachel; Yin, Fang-Fang; Horton, Janet KObjective
To analyze and evaluate accuracy and efficiency of IGRT process for preoperative partial breast radiosurgery.Methods
Patients were initially setup with skin marks and 5 steps were performed: (1) Initial orthogonal 2D kV images, (2) pre-treatment 3D CBCT images, (3) verification orthogonal 2D kV images, (4) treatment including mid-treatment 2D kV images (for the final 15 patients only), and (5) post-treatment orthogonal 2D kV or 3D CBCT images. Patient position was corrected at each step to align the biopsy clip and to verify surrounding soft tissue positioning.Results
The mean combined vector magnitude shifts and standard deviations at the 5 imaging steps were (1) 0.96 ± 0.69, (2) 0.33 ± 0.40, (3) 0.05 ± 0.12, (4) 0.15 ± 0.17, and (5) 0.27 ± 0.24 in cm. The mean total IGRT time was 40.2 ± 13.2 minutes. Each step was shortened by 2 to 5 minutes with improvements implemented. Overall, improvements in the IGRT process reduced the mean total IGRT time by approximately 20 minutes. Clip visibility was improved by implementing oblique orthogonal images.Conclusion
Multiple imaging steps confirmed accurate patient positioning. Appropriate planning and imaging strategies improved the effectiveness and efficiency of the IGRT process for preoperative partial breast radiosurgery.Item Open Access Automatic Planning of Whole Breast Radiation Therapy Using Machine Learning Models.(Frontiers in Oncology, 2019-01) Sheng, Yang; Li, Taoran; Yoo, Sua; Yin, Fang-Fang; Blitzblau, Rachel; Horton, Janet K; Ge, Yaorong; Wu, Q JackiePurpose: To develop an automatic treatment planning system for whole breast radiation therapy (WBRT) based on two intensity-modulated tangential fields, enabling near-real-time planning. Methods and Materials: A total of 40 WBRT plans from a single institution were included in this study under IRB approval. Twenty WBRT plans, 10 with single energy (SE, 6MV) and 10 with mixed energy (ME, 6/15MV), were randomly selected as training dataset to develop the methodology for automatic planning. The rest 10 SE cases and 10 ME cases served as validation. The auto-planning process consists of three steps. First, an energy prediction model was developed to automate energy selection. This model establishes an anatomy-energy relationship based on principle component analysis (PCA) of the gray level histograms from training cases' digitally reconstructed radiographs (DRRs). Second, a random forest (RF) model generates an initial fluence map using the selected energies. Third, the balance of overall dose contribution throughout the breast tissue is realized by automatically selecting anchor points and applying centrality correction. The proposed method was tested on the validation dataset. Non-parametric equivalence test was performed for plan quality metrics using one-sided Wilcoxon Signed-Rank test. Results: For validation, the auto-planning system suggested same energy choices as clinical-plans in 19 out of 20 cases. The mean (standard deviation, SD) of percent target volume covered by 100% prescription dose was 82.5% (4.2%) for auto-plans, and 79.3% (4.8%) for clinical-plans (p > 0.999). Mean (SD) volume receiving 105% Rx were 95.2 cc (90.7 cc) for auto-plans and 83.9 cc (87.2 cc) for clinical-plans (p = 0.108). Optimization time for auto-plan was <20 s while clinical manual planning takes between 30 min and 4 h. Conclusions: We developed an automatic treatment planning system that generates WBRT plans with optimal energy selection, clinically comparable plan quality, and significant reduction in planning time, allowing for near-real-time planning.Item Open Access Clinical Experience With Machine Learning-Based Automated Treatment Planning for Whole Breast Radiation Therapy.(Advances in radiation oncology, 2021-03) Yoo, Sua; Sheng, Yang; Blitzblau, Rachel; McDuff, Susan; Champ, Colin; Morrison, Jay; O'Neill, Leigh; Catalano, Suzanne; Yin, Fang-Fang; Wu, Q JackiePurpose
The machine learning-based automated treatment planning (MLAP) tool has been developed and evaluated for breast radiation therapy planning at our institution. We implemented MLAP for patient treatment and assessed our clinical experience for its performance.Methods and materials
A total of 102 patients of breast or chest wall treatment plans were prospectively evaluated with institutional review board approval. A human planner executed MLAP to create an auto-plan via automation of fluence maps generation. If judged necessary, a planner further fine-tuned the fluence maps to reach a final plan. Planners recorded the time required for auto-planning and manual modification. Target (ie, breast or chest wall and nodes) coverage and dose homogeneity were compared between the auto-plan and final plan.Results
Cases without nodes (n = 71) showed negligible (<1%) differences for target coverage and dose homogeneity between the auto-plan and final plan. Cases with nodes (n = 31) also showed negligible difference for target coverage. However, mean ± standard deviation of volume receiving 105% of the prescribed dose and maximum dose were reduced from 43.0% ± 26.3% to 39.4% ± 23.7% and 119.7% ± 9.5% to 114.4% ± 8.8% from auto-plan to final plan, respectively, all with P ≤ .01 for cases with nodes (n = 31). Mean ± standard deviation time spent for auto-plans and additional fluence modification for final plans were 12.1 ± 9.3 and 13.1 ± 12.9 minutes, respectively, for cases without nodes, and 16.4 ± 9.7 and 26.4 ± 16.4 minutes, respectively, for cases with nodes.Conclusions
The MLAP tool has been successfully implemented for routine clinical practice and has significantly improved planning efficiency. Clinical experience indicates that auto-plans are sufficient for target coverage, but improvement is warranted to reduce high dose volume for cases with nodal irradiation. This study demonstrates the clinical implementation of auto-planning for patient treatment and the significant importance of integrating human experience and feedback to improve MLAP for better clinical translation.