Item Open Access Accuracy of Effective Dose Estimation Using Single and Double Badges and an Evaluation of Organ Dose and Image Quality in Thoracic MDCT Scans Through a Comparison of Bismuth Shields and a Global Reduction in Tube Current(2012) Januzis, Natalie AnnPurpose: (1) To benchmark the accuracy of effective dose equivalent (EDE) of the single- and double-badge methods (NRC 2002-06) using the commercially available radiation monitors in clinical settings, (2) to study the transmission properties of various shielding materials, (3) to evaluate the accuracy of film badge readings compared to a calibrated ion chamber, (4) to benchmark the accuracy of effective dose (ED) of the single- and double-badge methods (NRC 2002-06) using the MOSFET method, and (5) to investigate the organ dose and image quality in a thoracic MDCT scan under the following conditions: (a) tube current modulation (TCM) without a Bismuth shield, (b) TCM with a Bismuth shield, and (c) manually reduced tube current (RTC) with no Bismuth shield.
Methods and Materials: (Project 1): Radiation workers in interventional radiology and cardiac catheterization laboratory were provided with two monitors and asked to place one at the collar and the other underneath the lead apron. Two commercial radiation monitor vendors were used for the study; both vendors were accredited by the National Voluntary Laboratory Accreditation Program (NVLAP). Effective dose equivalent (EDE) was computed by single-badge and double-badge methods based on the NRC Publication 2002-06. Data were plotted EDE1 (single badge) vs. collar reading and EDE2 (double badge) vs. collar reading. Data on EDE2 vs. collar reading were fitted by linear regression and a new equation for the EDE2collar was derived for routine clinical EDE estimation.
(Project 2): The transmission properties of lead aprons and a thyroid shield were measured using a 6-cc ion chamber and electrometer. These measurements were taken on a GE VCT (64 slice) scanner at 80, 100, and 120 kVp. The different types of lead aprons studied included lead free, lightweight lead, and fully leaded.
(Project 3): The accuracy of film badges was evaluated by comparing the reported deep dose equivalent of the film badge readings to the ion chamber readings measured during the same exposure. The measurements were made on a Philips Standard Radiography unit (Duke North, Room H1) at 80, 100, and 120 kVp. Two badges were exposed with the ion chamber per energy.
(Project 4): An adult male anthropomorphic phantom was loaded with 20 diagnostic MOSFET detectors and scanned without lead aprons using a whole body computed tomography (CT) protocol. All measurements were taken on a GE VCT (64 slice) scanner at 80, 100, 120 kVp. Two commercial film badges were placed on the phantom at the collar location and waist location. Individual organ doses in the phantom were corrected for lead apron attenuation factor and ED was computed using ICRP 103 tissue weighting factors. The single badge conversion coefficient (CC) was determined for each energy by taking the ratio of the ED to collar badge reading. The reported deep dose equivalent for the collar badge was plotted against the MOSFET effective dose and a new equation for EDcollar was derived.
(Project 5): Organ dose was measured with MOSFETs using an adult female anthropomorphic phantom; the phantom was scanned with pulmonary embolus protocol. All measurements were performed with a 64-slice scanner at 120 kVp. The reference exposure and reduced exposure (with 4-ply Bi shield) was measured with an ion chamber located at the level of the breast. The tube current was reduced by normalizing the reference tube current to the ratio of the reduced exposure to the reference exposure. Image quality was measured using a high contrast insert placed in the lung. Regions of interest (ROIs) were drawn in the breast, lung, and heart to measure HU change and noise. ROIs were drawn in the lung and high contrast insert to measure signal-to-noise ratio (SNR) and percent contrast (%Contrast).
Results: (Project 1): From the data, it can be seen that EDE1 read about a factor of six greater than EDE2. The new equation for EDE2collar yielded a slope of 0.06992, a y-intercept of -1.682, and a r2 value of 0.9081.
(Project 2): The transmission for the fully leaded, lightweight lead aprons, and lead free apron were 3.19%, 3.71%, and 7.06% at 80 kVp; 6.58%, 8.07%, and 13.04% at 100 kVp; 7.61%, 12.05%, and 17.84% at 120 kVp, respectively. The attenuation for the thyroid shield was 3.02%, 6.35%, and 7.74% at 80, 100, and 120 kVp, respectively.
(Project 3): The average badge reading was 3.49 ± 1.01% mSv at 80 kVp; 4.80 ± 7.37% mSv at 100 kVp; 4.90 ± 14.1% mSv at 120 kVp. The dose to soft tissue measured by the ion chamber was 4.53 mSv at 80 kVp; 5.71 mSv at 100 kVp; 6.35 mSv at 120 kVp. The film badge reading differed from the ion chamber measurement by -22.8%, -15.9%, and -22.9% at 80, 100, and 120 kVp, respectively.
(Project 4): The ED and % difference between the single-badge method (NRC 2002-06) and the MOSFET method were as follows: 11.65 mSv vs. 0.50 mSv (2331%) for 80 kVp; 27.85 mSv vs. 2.14mSv (1301%) for 100 kVp; 38.59 mSv vs. 4.98 mSv (775%) for 120 kVp, respectively. The ED and % difference between the double-badge method (NRC 2002-06) and the MOSFET method were as follows: 4.07 mSv vs. 0.50 mSv (808%) for 80 kVp; 16.9 mSv vs. 2.14 mSv (791%) for 100 kVp; 25.4 mSv vs. 4.98 mSv (510%) for 120 kVp, respectively. The single badge conversion factors were 0.01 ± 14.8% (80 kVp), 0.02 ± 9.5% (100 kVp), and 0.04 ± 15.7% (120 kVp). The plot of collar badge reading vs. MOSFET effective dose yielded an equation with a slope of 0.0483, a y-intercept of -1.6517, and a R2 value of 0.92929.
(Project 5): Organ doses (mGy) for the three scans (TCM, TCM with Bi, and RTC with no Bi) were 45.8, 27.1, and 27.8 to the breast; 51.6, 47.0, and 29.1 to the lung; and 42.1, 35.0 and 24.9 to the heart, respectively. HU increase was greatest in the TCM with Bi scan. The SNRs were 77.1, 63.7, and 59.2 and the %Contrast values were 369.5, 347.1, and 362.7 with TCM, TCM with Bi, and RTC, respectively.
Conclusions: (Project 1): A new EDE estimation method has been developed based on the results of two-badge system. The method would enable us to compute new EDE values knowing only the collar badge reading. Since EDE2 reads a factor of six less than EDE1, this provides a realistic advantage in regulatory compliance for interventional and cardiac catheterization personnel. Further, new EDE conversion coefficients should be developed for better assessment of EDE.
(Project 2): The fully leaded shielding materials had the lowest percent transmission. It should be noted that radiation workers are generally exposed to only scattered radiation of lower energy. Although this study did not measure attenuation properties at lower energies, it is expected that the percentage of attenuation will only increase with lower energies.
(Project 3): The reported deep dose equivalent (DDE) underestimated the dose to soft tissue compared to the calibrated ion chamber readings. This may be due to the fact that DDE is the dose equivalent at a depth of 10 mm.
(Project 4): Current regulatory ED conversion coefficient (CC) with single collar badge is 0.3; for double-badge system, they are 0.04 and 1.5 for the collar and under the apron respectively. Based on our findings we recommend the current collar CC be dropped due to the overestimation of ED. Since occupational workers are exposed mainly to scattered x-rays of lower energy, a collar CC of 0.01 (80 kVp data) may be a more viable option. The double badge system seems to provide a better coefficient for the collar as 0.04; however, exposure readings under the apron are usually negligible to zero with lead aprons.
(Project 5): For thoracic CT using RTC will result in similar global reduction in organ dose; the use of Bismuth with TCM will lead to an overall decrease in organ dose and more marked dose reduction for the breast. There was a significant difference in SNR (p = 0.0003) and %Contrast (p < 0.0001) in the TCM with Bismuth scan compared to the reference scan (TCM). The RTC scan also demonstrated a significant decrease in SNR and %Contrast with p < 0.0001 for both. While the TCM scan demonstrated superior image quality, the trade-off is in the increased dose to the breast.
Item Open Access Radiation Dose to the Lens of the Eye from Computed Tomography Scans of the Head(2016) Januzis, Natalie AnnWhile it is well known that exposure to radiation can result in cataract formation, questions still remain about the presence of a dose threshold in radiation cataractogenesis. Since the exposure history from diagnostic CT exams is well documented in a patient’s medical record, the population of patients chronically exposed to radiation from head CT exams may be an interesting area to explore for further research in this area. However, there are some challenges in estimating lens dose from head CT exams. An accurate lens dosimetry model would have to account for differences in imaging protocols, differences in head size, and the use of any dose reduction methods.
The overall objective of this dissertation was to develop a comprehensive method to estimate radiation dose to the lens of the eye for patients receiving CT scans of the head. This research is comprised of a physics component, in which a lens dosimetry model was derived for head CT, and a clinical component, which involved the application of that dosimetry model to patient data.
The physics component includes experiments related to the physical measurement of the radiation dose to the lens by various types of dosimeters placed within anthropomorphic phantoms. These dosimeters include high-sensitivity MOSFETs, TLDs, and radiochromic film. The six anthropomorphic phantoms used in these experiments range in age from newborn to adult.
First, the lens dose from five clinically relevant head CT protocols was measured in the anthropomorphic phantoms with MOSFET dosimeters on two state-of-the-art CT scanners. The volume CT dose index (CTDIvol), which is a standard CT output index, was compared to the measured lens doses. Phantom age-specific CTDIvol-to-lens dose conversion factors were derived using linear regression analysis. Since head size can vary among individuals of the same age, a method was derived to estimate the CTDIvol-to-lens dose conversion factor using the effective head diameter. These conversion factors were derived for each scanner individually, but also were derived with the combined data from the two scanners as a means to investigate the feasibility of a scanner-independent method. Using the scanner-independent method to derive the CTDIvol-to-lens dose conversion factor from the effective head diameter, most of the fitted lens dose values fell within 10-15% of the measured values from the phantom study, suggesting that this is a fairly accurate method of estimating lens dose from the CTDIvol with knowledge of the patient’s head size.
Second, the dose reduction potential of organ-based tube current modulation (OB-TCM) and its effect on the CTDIvol-to-lens dose estimation method was investigated. The lens dose was measured with MOSFET dosimeters placed within the same six anthropomorphic phantoms. The phantoms were scanned with the five clinical head CT protocols with OB-TCM enabled on the one scanner model at our institution equipped with this software. The average decrease in lens dose with OB-TCM ranged from 13.5 to 26.0%. Using the size-specific method to derive the CTDIvol-to-lens dose conversion factor from the effective head diameter for protocols with OB-TCM, the majority of the fitted lens dose values fell within 15-18% of the measured values from the phantom study.
Third, the effect of gantry angulation on lens dose was investigated by measuring the lens dose with TLDs placed within the six anthropomorphic phantoms. The 2-dimensional spatial distribution of dose within the areas of the phantoms containing the orbit was measured with radiochromic film. A method was derived to determine the CTDIvol-to-lens dose conversion factor based upon distance from the primary beam scan range to the lens. The average dose to the lens region decreased substantially for almost all the phantoms (ranging from 67 to 92%) when the orbit was exposed to scattered radiation compared to the primary beam. The effectiveness of this method to reduce lens dose is highly dependent upon the shape and size of the head, which influences whether or not the angled scan range coverage can include the entire brain volume and still avoid the orbit.
The clinical component of this dissertation involved performing retrospective patient studies in the pediatric and adult populations, and reconstructing the lens doses from head CT examinations with the methods derived in the physics component. The cumulative lens doses in the patients selected for the retrospective study ranged from 40 to 1020 mGy in the pediatric group, and 53 to 2900 mGy in the adult group.
This dissertation represents a comprehensive approach to lens of the eye dosimetry in CT imaging of the head. The collected data and derived formulas can be used in future studies on radiation-induced cataracts from repeated CT imaging of the head. Additionally, it can be used in the areas of personalized patient dose management, and protocol optimization and clinician training.