Item Open Access Association Between Comorbidities and Outcomes in Heart Failure Patients With and Without an Implantable Cardioverter-Defibrillator for Primary Prevention.(J Am Heart Assoc, 2015-08-06) Khazanie, Prateeti; Hellkamp, Anne S; Fonarow, Gregg C; Bhatt, Deepak L; Masoudi, Frederick A; Anstrom, Kevin J; Heidenreich, Paul A; Yancy, Clyde W; Curtis, Lesley H; Hernandez, Adrian F; Peterson, Eric D; Al-Khatib, Sana MBACKGROUND: Implantable cardioverter-defibrillator (ICD) therapy is associated with improved outcomes in patients with heart failure (HF), but whether this association holds among older patients with multiple comorbid illnesses and worse HF burden remains unclear. METHODS AND RESULTS: Using the National Cardiovascular Data Registry's ICD Registry and the Get With The Guidelines-Heart Failure (GWTG-HF) registry linked with Medicare claims, we examined outcomes associated with primary-prevention ICD versus no ICD among HF patients aged ≥65 years in clinical practice. We included patients with an ejection fraction ≤35% who received (ICD Registry) and who did not receive (GWTG-HF) an ICD. Compared with patients with an ICD, patients in the non-ICD group were older and more likely to be female and white. In matched cohorts, the 3-year adjusted mortality rate was lower in the ICD group versus the non-ICD group (46.7% versus 55.8%; adjusted hazard ratio [HR] 0.76; 95% CI 0.69 to 0.83). There was no associated difference in all-cause readmission (HR 0.99; 95% CI 0.92 to 1.08) but a lower risk of HF readmission (HR 0.88; 95% CI 0.80 to 0.97). When compared with no ICD, ICDs were also associated with better survival in patients with ≤3 comorbidities (HR 0.77; 95% CI 0.69 to 0.87) and >3 comorbidities (HR 0.77; 95% CI 0.64 to 0.93) and in patients with no hospitalization for HF (HR 0.75; 95% CI 0.65 to 0.86) and at least 1 prior HF hospitalization (HR 0.69; 95% CI 0.58 to 0.82). In subgroup analyses, there were no interactions between ICD and mortality risk for comorbidity burden (P=0.95) and for prior HF hospitalization (P=0.46). CONCLUSION: Among older HF patients, ICDs for primary prevention were associated with lower risk of mortality even among those with high comorbid illness burden and prior HF hospitalization.Item Open Access Cable externalization and electrical failure of the Riata family of implantable cardioverter-defibrillator leads: A systematic review and meta-analysis.(Heart rhythm, 2015-06) Zeitler, Emily P; Pokorney, Sean D; Zhou, Ke; Lewis, Robert K; Greenfield, Ruth Ann; Daubert, James P; Matchar, David B; Piccini, Jonathan PBackground
The Riata class of defibrillator leads were placed under US Food and Drug Association (FDA) advisory as of November 2011 because of high rates of cable externalization (CE) and electrical failure (EF). The overall rates of these complications remain unknown.Objective
The purpose of this study was to systematically search the literature for rates of Riata lead failure and to perform a meta-analysis to estimate failure rates.Methods
We conducted a meta-analysis of observational studies examining the rates of EF, CE, and the interaction of the two. We identified 23 English language manuscripts addressing 1 or more of these questions.Results
Across 23 studies, the overall CE rate was 23.1% (95% confidence interval [CI] 19.0%-27.6%). The overall EF rate was 6.3% (95% CI 4.7%-8.2%). The presence of CE was associated with a more than 6-fold increase in the rate of EF compared to no CE (17.3% [95% CI 11.2%-25.9%] vs 2.7% [95% CI 1.4%-5.2%], respectively). The rate of CE was 3-fold higher for 8Fr leads compared to 7Fr leads, but rates of EF were similar (4.6%; 95% CI 3.2-6.6] and 3.9%; 95% CI 2.4-6.1], respectively). Rates of both CE and EF were higher in dual coil vs single coil leads, but confidence intervals overlapped.Conclusion
In clinical practice, rates of CE in Riata leads are substantial. While CE is associated with a significant increase in the risk of EF, the incidence of EF without externalization is not trivial.Item Open Access Follow-up of patients with new cardiovascular implantable electronic devices: are experts' recommendations implemented in routine clinical practice?(Circulation. Arrhythmia and electrophysiology, 2013-02) Al-Khatib, Sana M; Mi, Xiaojuan; Wilkoff, Bruce L; Qualls, Laura G; Frazier-Mills, Camille; Setoguchi, Soko; Hess, Paul L; Curtis, Lesley HBackground
A 2008 expert consensus statement outlined the minimum frequency of follow-up of patients with cardiovascular implantable electronic devices (CIEDs).Methods and results
We studied 38 055 Medicare beneficiaries who received a new CIED between January 1, 2005, and June 30, 2009. The main outcome measure was variation of follow-up by patient factors and year of device implantation. We determined the number of patients who were eligible for and attended an in-person CIED follow-up visit within 2 to 12 weeks, 0 to 16 weeks, and 1 year after implantation. Among eligible patients, 42.4% had an initial in-person visit within 2 to 12 weeks. This visit was significantly more common among white patients than black patients and patients of other races (43.0% versus 36.8% versus 40.5%; P<0.001). Follow-up within 2 to 12 weeks improved from 40.3% in 2005 to 55.1% in 2009 (P<0.001 for trend). The rate of follow-up within 0 to 16 weeks was 65.1% and improved considerably from 2005 to 2009 (62.3%-79.6%; P<0.001 for trend). Within 1 year, 78.0% of the overall population had at least 1 in-person CIED follow-up visit.Conclusions
Although most Medicare beneficiaries who received a new CIED between 2005 and 2009 did not have an initial in-person CIED follow-up visit within 2 to 12 weeks after device implantation, the rate of initial follow-up improved appreciably over time. This CIED follow-up visit was significantly more common in white patients than in patients of other races.Item Open Access Implant healing in experimental animal models of diabetes.(J Diabetes Sci Technol, 2011-05-01) Le, NN; Rose, MB; Levinson, H; Klitzman, BDiabetes mellitus is becoming increasingly prevalent worldwide. Additionally, there is an increasing number of patients receiving implantable devices such as glucose sensors and orthopedic implants. Thus, it is likely that the number of diabetic patients receiving these devices will also increase. Even though implantable medical devices are considered biocompatible by the Food and Drug Administration, the adverse tissue healing that occurs adjacent to these foreign objects is a leading cause of their failure. This foreign body response leads to fibrosis, encapsulation of the device, and a reduction or cessation of device performance. A second adverse event is microbial infection of implanted devices, which can lead to persistent local and systemic infections and also exacerbates the fibrotic response. Nearly half of all nosocomial infections are associated with the presence of an indwelling medical device. Events associated with both the foreign body response and implant infection can necessitate device removal and may lead to amputation, which is associated with significant morbidity and cost. Diabetes mellitus is generally indicated as a risk factor for the infection of a variety of implants such as prosthetic joints, pacemakers, implantable cardioverter defibrillators, penile implants, and urinary catheters. Implant infection rates in diabetic patients vary depending upon the implant and the microorganism, however, for example, diabetes was found to be a significant variable associated with a nearly 7.2% infection rate for implantable cardioverter defibrillators by the microorganism Candida albicans. While research has elucidated many of the altered mechanisms of diabetic cutaneous wound healing, the internal healing adjacent to indwelling medical devices in a diabetic model has rarely been studied. Understanding this healing process is crucial to facilitating improved device design. The purpose of this article is to summarize the physiologic factors that influence wound healing and infection in diabetic patients, to review research concerning diabetes and biomedical implants and device infection, and to critically analyze which diabetic animal model might be advantageous for assessing internal healing adjacent to implanted devices.Item Open Access Isolated Noncompaction of the Left Ventricle in Adults.(J Am Coll Cardiol, 2015-08-04) Hussein, Ayman; Karimianpour, Ahmadreza; Collier, Patrick; Krasuski, Richard AIsolated left ventricular noncompaction (ILVNC) is a cardiomyopathy that was first described in 1926 as a "spongy myocardium." The disorder results from intrauterine arrest of compaction of the loose interwoven meshwork of the fetal myocardial primordium and subsequent persistence of deep trabecular recesses in the myocardial wall. The classical clinical presentation is a triad of heart failure, arrhythmias, and embolic events from mural thrombi. ILVNC has been associated with several autosomal dominant, X-linked, and mitochondrial genetic mutations that are also shared among other cardiomyopathies. Over the past decade, ILVNC has been subject to intensive research, as it increases the risk for sudden cardiac death. This review focuses on the current understanding of ILVNC in adult populations and attempts to provide organized insight into the disease process, screening, diagnosis, management, role of device therapy, and prognosis.Item Open Access Optimal management of Riata leads with no known electrical abnormalities or externalization: a decision analysis.(Journal of cardiovascular electrophysiology, 2015-02) Pokorney, Sean D; Zhou, Ke; Matchar, David B; Love, Sean; Zeitler, Emily P; Lewis, Robert; Piccini, Jonathan PIntroduction
Riata and Riata ST implantable cardioverter-defibrillator (ICD) leads (St. Jude Medical, Sylmar, CA, USA) can develop conductor cable externalization and/or electrical failure. Optimal management of these leads remains unknown.Methods and results
A Markov model compared 4 lead management strategies: (1) routine device interrogation for electrical failure, (2) systematic yearly fluoroscopic screening and routine device interrogation, (3) implantation of new ICD lead with capping of the in situ lead, and (4) implantation of new ICD lead with extraction of the in situ lead. The base case was a 64-year-old primary prevention ICD patient. Modeling demonstrated average life expectancies as follows: capping with new lead implanted at 134.5 months, extraction with new lead implanted at 134.0 months, fluoroscopy with routine interrogation at 133.9 months, and routine interrogation at 133.5 months. One-way sensitivity analyses identified capping as the preferred strategy with only one parameter having a threshold value: when risk of nonarrhythmic death associated with lead abandonment is greater than 0.05% per year, lead extraction is preferred over capping. A second-order Monte Carlo simulation (n = 10,000), as a probabilistic sensitivity analysis, found that lead revision was favored with 100% certainty (extraction 76% and capping 24%).Conclusions
Overall there were minimal differences in survival with monitoring versus active lead management approaches. There is no evidence to support fluoroscopic screening for externalization of Riata or Riata ST leads.Item Open Access Use of cardiac implantable electronic devices in older adults with cognitive impairment.(JAMA Intern Med, 2014-09) Fowler, Nicole R; Johnson, Kim G; Li, Jie; Moore, Charity G; Saba, Samir; Lopez, Oscar L; Barnato, Amber E