Evaluation of Volumetric Losses During Radiation Therapy Using Image Guidance of Electronic Portal Imaging Device

Thumbnail
View / Download
10.1 Mb
Date
2010
Author
Senick, Scott Michael
Advisor
Chang, Zheng
Repository Usage Stats
393
views
239
downloads
Abstract

<bold>Purpose:</bold> Changes in patient volume, due to tumor shrinkage, dehydration, dysphagia and atrophy, could present issues in the accuracy of dosimetry throughout the course of treatment. The aim of this work is to study the dosimetric impacts of the volumetric changes during IMRT and to investigate the feasibilities of electronic portal imaging device (EPID) in predicting the impacts. Materials and <bold>Methods:</bold> An anthropomorphic head and neck phantom was used to represent two scenarios: symmetric and asymmetric volume loss. The phantom was simulated and planned according to the head and neck protocols used in our clinic. Dose volume histograms (DVH) were generated for each set up scenario and were used to calculate the integral dose expected at the coincident volume of the phantom. During treatment delivery, the EPID captured exit fluence of each beam at each level of bolus thickness. These images were quantitatively analyzed using gamma analysis with criteria of 3% and 3mm dose difference and distance-to-agreement respectively. <bold>Results:</bold> Comparing maximum to minimum volume in the symmetric situation with DVH generated in Eclipse show substantial fluctuations in dose. When comparing five layers of bolus material to zero layers of bolus material, the changes were most significant. The asymmetric volume change predicted dose fluctuations that were less significant than the symmetric phantom. As for gamma analysis, a quantitative evaluation of the integrated dose fluence, captured by the EPID, showed extreme variability in the images with five layers of bolus when compared to images with no bolus. Less significant variation was shown in layers of closer thicknesses, as expected. <bold>Conclusions:</bold> The phantom study indicates that volume loss could contribute to clinically considerable changes in the dose delivered to target and organs at risk. The proposed technique using EPID could provide valuable information about the variation of dose due to volumetric changes and might be potentially useful.

Type
Master's thesis
Department
Medical Physics
Subject
Physics, Radiation
Health Sciences, Oncology
EPID
Portal Imager evaluation
Volumetric Phantom Study
Permalink
https://hdl.handle.net/10161/3062
Citation
Senick, Scott Michael (2010). Evaluation of Volumetric Losses During Radiation Therapy Using Image Guidance of Electronic Portal Imaging Device. Master's thesis, Duke University. Retrieved from https://hdl.handle.net/10161/3062.
Collections
  • Masters Theses
More Info
Show full item record
Creative Commons License
This work is licensed under a Creative Commons Attribution-Noncommercial-No Derivative Works 3.0 United States License.

Rights for Collection: Masters Theses


Works are deposited here by their authors, and represent their research and opinions, not that of Duke University. Some materials and descriptions may include offensive content. More info

Make Your Work Available Here

How to Deposit

Browse

All of DukeSpaceCommunities & CollectionsAuthorsTitlesTypesBy Issue DateDepartmentsAffiliations of Duke Author(s)SubjectsBy Submit DateThis CollectionAuthorsTitlesTypesBy Issue DateDepartmentsAffiliations of Duke Author(s)SubjectsBy Submit Date

My Account

LoginRegister

Statistics

View Usage Statistics
Duke University Libraries

Contact Us

411 Chapel Drive
Durham, NC 27708
(919) 660-5870
Perkins Library Service Desk

Digital Repositories at Duke

  • Report a problem with the repositories
  • About digital repositories at Duke
  • Accessibility Policy
  • Deaccession and DMCA Takedown Policy

TwitterFacebookYouTubeFlickrInstagramBlogs

Sign Up for Our Newsletter
  • Re-use & Attribution / Privacy
  • Harmful Language Statement
  • Support the Libraries
Duke University