Characterization of porous, dexamethasone-releasing polyurethane coatings for glucose sensors.

dc.contributor.author

Vallejo-Heligon, Suzana G

dc.contributor.author

Klitzman, Bruce

dc.contributor.author

Reichert, William M

dc.coverage.spatial

England

dc.date.accessioned

2014-10-20T15:48:29Z

dc.date.issued

2014-11

dc.description.abstract

Commercially available implantable needle-type glucose sensors for diabetes management are robust analytically but can be unreliable clinically primarily due to tissue-sensor interactions. Here, we present the physical, drug release and bioactivity characterization of tubular, porous dexamethasone (Dex)-releasing polyurethane coatings designed to attenuate local inflammation at the tissue-sensor interface. Porous polyurethane coatings were produced by the salt-leaching/gas-foaming method. Scanning electron microscopy and micro-computed tomography (micro-CT) showed controlled porosity and coating thickness. In vitro drug release from coatings monitored over 2 weeks presented an initial fast release followed by a slower release. Total release from coatings was highly dependent on initial drug loading amount. Functional in vitro testing of glucose sensors deployed with porous coatings against glucose standards demonstrated that highly porous coatings minimally affected signal strength and response rate. Bioactivity of the released drug was determined by monitoring Dex-mediated, dose-dependent apoptosis of human peripheral blood derived monocytes in culture. Acute animal studies were used to determine the appropriate Dex payload for the implanted porous coatings. Pilot short-term animal studies showed that Dex released from porous coatings implanted in rat subcutis attenuated the initial inflammatory response to sensor implantation. These results suggest that deploying sensors with the porous, Dex-releasing coatings is a promising strategy to improve glucose sensor performance.

dc.identifier

http://www.ncbi.nlm.nih.gov/pubmed/25065548

dc.identifier

S1742-7061(14)00321-3

dc.identifier.eissn

1878-7568

dc.identifier.uri

https://hdl.handle.net/10161/9194

dc.language

eng

dc.relation.ispartof

Acta Biomaterialia

dc.relation.isversionof

10.1016/j.actbio.2014.07.019

dc.subject

Biosensor

dc.subject

Controlled drug release

dc.subject

Foreign body response

dc.subject

Inflammation

dc.subject

Porosity

dc.subject

Animals

dc.subject

Anti-Inflammatory Agents

dc.subject

Biosensing Techniques

dc.subject

Calorimetry, Differential Scanning

dc.subject

Coated Materials, Biocompatible

dc.subject

Dexamethasone

dc.subject

Glucose

dc.subject

Humans

dc.subject

Pilot Projects

dc.subject

Polyurethanes

dc.subject

Porosity

dc.subject

Prosthesis Implantation

dc.subject

Rats, Sprague-Dawley

dc.subject

Time Factors

dc.subject

X-Ray Microtomography

dc.title

Characterization of porous, dexamethasone-releasing polyurethane coatings for glucose sensors.

dc.type

Journal article

pubs.author-url

http://www.ncbi.nlm.nih.gov/pubmed/25065548

pubs.begin-page

4629

pubs.end-page

4638

pubs.issue

11

pubs.organisational-group

Basic Science Departments

pubs.organisational-group

Biomedical Engineering

pubs.organisational-group

Cell Biology

pubs.organisational-group

Clinical Science Departments

pubs.organisational-group

Duke

pubs.organisational-group

Pratt School of Engineering

pubs.organisational-group

School of Medicine

pubs.organisational-group

Surgery

pubs.organisational-group

Surgery, Plastic, Maxillofacial, and Oral Surgery

pubs.publication-status

Published

pubs.volume

10

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Vallejo-Heligon-DexPU-ActaBiomat2014.pdf
Size:
2.17 MB
Format:
Adobe Portable Document Format
Description:
Published version