Patient-Informed Organ Dose Estimation in Clinical CT: Implementation and Effective Dose Assessment in 1048 Clinical Patients.

dc.contributor.author

Fu, Wanyi

dc.contributor.author

Ria, Francesco

dc.contributor.author

Segars, William Paul

dc.contributor.author

Choudhury, Kingshuk Roy

dc.contributor.author

Wilson, Joshua M

dc.contributor.author

Kapadia, Anuj J

dc.contributor.author

Samei, Ehsan

dc.date.accessioned

2021-01-25T20:41:20Z

dc.date.available

2021-01-25T20:41:20Z

dc.date.issued

2021-01-21

dc.date.updated

2021-01-25T20:40:35Z

dc.description.abstract

OBJECTIVE. The purpose of this study is to comprehensively implement a patient-informed organ dose monitoring framework for clinical CT and compare the effective dose (ED) according to the patient-informed organ dose with ED according to the dose-length product (DLP) in 1048 patients. MATERIALS AND METHODS. Organ doses for a given examination are computed by matching the topogram to a computational phantom from a library of anthropomorphic phantoms and scaling the fixed tube current dose coefficients by the examination volume CT dose index (CTDIvol) and the tube-current modulation using a previously validated convolution-based technique. In this study, the library was expanded to 58 adult, 56 pediatric, five pregnant, and 12 International Commission on Radiological Protection (ICRP) reference models, and the technique was extended to include multiple protocols, a bias correction, and uncertainty estimates. The method was implemented in a clinical monitoring system to estimate organ dose and organ dose-based ED for 647 abdomen-pelvis and 401 chest examinations, which were compared with DLP-based ED using a t test. RESULTS. For the majority of the organs, the maximum errors in organ dose estimation were 18% and 8%, averaged across all protocols, without and with bias correction, respectively. For the patient examinations, DLP-based ED was significantly different from organ dose-based ED by as much as 190.9% and 234.7% for chest and abdomen-pelvis scans, respectively (mean, 9.0% and 24.3%). The differences were statistically significant (p < .001) and exhibited overestimation for larger-sized patients and underestimation for smaller-sized patients. CONCLUSION. A patient-informed organ dose estimation framework was comprehensively implemented applicable to clinical imaging of adult, pediatric, and pregnant patients. Compared with organ dose-based ED, DLP-based ED may overestimate effective dose for larger-sized patients and underestimate it for smaller-sized patients.

dc.identifier.issn

0361-803X

dc.identifier.issn

1546-3141

dc.identifier.uri

https://hdl.handle.net/10161/22242

dc.language

eng

dc.publisher

American Roentgen Ray Society

dc.relation.ispartof

AJR. American journal of roentgenology

dc.relation.isversionof

10.2214/ajr.19.22482

dc.subject

CT

dc.subject

effective dose

dc.subject

patient-specific organ dose

dc.subject

radiation burden characterization

dc.title

Patient-Informed Organ Dose Estimation in Clinical CT: Implementation and Effective Dose Assessment in 1048 Clinical Patients.

dc.type

Journal article

duke.contributor.orcid

Ria, Francesco|0000-0001-5902-7396

duke.contributor.orcid

Wilson, Joshua M|0000-0002-4175-6301

duke.contributor.orcid

Samei, Ehsan|0000-0001-7451-3309

pubs.begin-page

1

pubs.end-page

11

pubs.organisational-group

Staff

pubs.organisational-group

Duke

pubs.publication-status

Published

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Fu_et_al_ajr.19.22482.pdf
Size:
1.78 MB
Format:
Adobe Portable Document Format
Description:
Published version