Flattening filter-free accelerators: a report from the AAPM Therapy Emerging Technology Assessment Work Group.

dc.contributor.author

Xiao, Ying

dc.contributor.author

Kry, Stephen F

dc.contributor.author

Popple, Richard

dc.contributor.author

Yorke, Ellen

dc.contributor.author

Papanikolaou, Niko

dc.contributor.author

Stathakis, Sotirios

dc.contributor.author

Xia, Ping

dc.contributor.author

Huq, Saiful

dc.contributor.author

Bayouth, John

dc.contributor.author

Galvin, James

dc.contributor.author

Yin, Fang-Fang

dc.date.accessioned

2019-10-04T16:18:19Z

dc.date.available

2019-10-04T16:18:19Z

dc.date.issued

2015-05-08

dc.date.updated

2019-10-04T16:18:17Z

dc.description.abstract

This report describes the current state of flattening filter-free (FFF) radiotherapy beams implemented on conventional linear accelerators, and is aimed primarily at practicing medical physicists. The Therapy Emerging Technology Assessment Work Group of the American Association of Physicists in Medicine (AAPM) formed a writing group to assess FFF technology. The published literature on FFF technology was reviewed, along with technical specifications provided by vendors. Based on this information, supplemented by the clinical experience of the group members, consensus guidelines and recommendations for implementation of FFF technology were developed. Areas in need of further investigation were identified. Removing the flattening filter increases beam intensity, especially near the central axis. Increased intensity reduces treatment time, especially for high-dose stereotactic radiotherapy/radiosurgery (SRT/SRS). Furthermore, removing the flattening filter reduces out-of-field dose and improves beam modeling accuracy. FFF beams are advantageous for small field (e.g., SRS) treatments and are appropriate for intensity-modulated radiotherapy (IMRT). For conventional 3D radiotherapy of large targets, FFF beams may be disadvantageous compared to flattened beams because of the heterogeneity of FFF beam across the target (unless modulation is employed). For any application, the nonflat beam characteristics and substantially higher dose rates require consideration during the commissioning and quality assurance processes relative to flattened beams, and the appropriate clinical use of the technology needs to be identified. Consideration also needs to be given to these unique characteristics when undertaking facility planning. Several areas still warrant further research and development. Recommendations pertinent to FFF technology, including acceptance testing, commissioning, quality assurance, radiation safety, and facility planning, are presented. Examples of clinical applications are provided. Several of the areas in which future research and development are needed are also indicated.

dc.identifier.issn

1526-9914

dc.identifier.issn

1526-9914

dc.identifier.uri

https://hdl.handle.net/10161/19406

dc.language

eng

dc.publisher

Wiley

dc.relation.ispartof

Journal of applied clinical medical physics

dc.relation.isversionof

10.1120/jacmp.v16i3.5219

dc.subject

Radiotherapy, Conformal

dc.subject

Filtration

dc.subject

Equipment Design

dc.subject

Equipment Failure Analysis

dc.subject

Health Physics

dc.subject

Radiation Protection

dc.subject

Particle Accelerators

dc.subject

Technology Assessment, Biomedical

dc.subject

United States

dc.subject

Practice Guidelines as Topic

dc.title

Flattening filter-free accelerators: a report from the AAPM Therapy Emerging Technology Assessment Work Group.

dc.type

Journal article

duke.contributor.orcid

Yin, Fang-Fang|0000-0002-2025-4740

pubs.begin-page

5219

pubs.issue

3

pubs.organisational-group

School of Medicine

pubs.organisational-group

Duke

pubs.organisational-group

Duke Kunshan University Faculty

pubs.organisational-group

Duke Kunshan University

pubs.organisational-group

Duke Cancer Institute

pubs.organisational-group

Institutes and Centers

pubs.organisational-group

Radiation Oncology

pubs.organisational-group

Clinical Science Departments

pubs.publication-status

Published

pubs.volume

16

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Flattening filter-free accelerators: a report from the AAPM Therapy Emerging Technology Assessment Work Group.pdf
Size:
520.26 KB
Format:
Adobe Portable Document Format