Electrically Conductive Silicone Interface For Myoelectric Prostheses With Silicone Socket Suspension

Loading...

Date

2011

Journal Title

Journal ISSN

Volume Title

Repository Usage Stats

555
views
192
downloads

Abstract

Silicone socket suspension technology for the upper extremity amputee has been proven to provide increased suspension and range of motion over conventional self suspending sockets. Application of silicone socket suspension has greatly benefitted clients with very short residual limbs and disarticulation level amputations. However, the use of silicone suspension sockets with myoelectric control has presented some problems, namely relating to the wear and tear on electrical control cables and connections. Managing the interface between silicone socket and hard socket without compromise to signal loss via an electrical-mechanical interface about the pin lock has been investigated, and although function was acceptable, this approach was technically complex. An alternate and potentially simple solution is the use of an electrically conductive silicone interface within the silicone socket and localized at the electrode site. To investigate the feasibility of this approach, a prosthetic socket and silicon sleeve using the conductive material was fabricated and evaluated on a single subject. Signal quality was found to be acceptable but further work is needed to assess the factors that can be targeted to further improve the signal-to-noise ratio. This approach has the potential to reduce the technical requirements in achieving usable EMG signal capture.

Type

Department

Description

Provenance

Subjects

Citation

Proceedings of the MEC'11 conference, UNB; 2011.

Copyright 2002, 2005 and 2008, The University of New Brunswick.

This work is licensed under a Creative Commons Attribution-Noncommercial-No Derivative Works 3.0 United States License. Creative Commons License