dc.description.abstract

The maturation of photon-counting detector (PCD) technology promises to enhance routine CT imaging applications with high-fidelity spectral information. In this paper, we demonstrate the power of this synergy and our complementary reconstruction techniques, performing 4D, cardiac PCD-CT data acquisition and reconstruction in a mouse model of atherosclerosis, including calcified plaque. Specifically, in vivo cardiac micro-CT scans were performed in four ApoE knockout mice, following their development of calcified plaques. The scans were performed with a prototype PCD (DECTRIS, Ltd.) with 4 energy thresholds. Projections were sampled every 10 ms with a 10 ms exposure, allowing the reconstruction of 10 cardiac phases at each of 4 energies (40 total 3D volumes per mouse scan). Reconstruction was performed iteratively using the split Bregman method with constraints on spectral rank and spatio-temporal gradient sparsity. The reconstructed images represent the first in vivo, 4D PCD-CT data in a mouse model of atherosclerosis. Robust regularization during iterative reconstruction yields high-fidelity results: an 8-fold reduction in noise standard deviation for the highest energy threshold (relative to unregularized algebraic reconstruction), while absolute spectral bias measurements remain below 13 Hounsfield units across all energy thresholds and scans. Qualitatively, image domain material decomposition results show clear separation of iodinated contrast and soft tissue from calcified plaque in the in vivo data. Quantitatively, spatial, spectral, and temporal fidelity are verified through a water phantom scan and a realistic MOBY phantom simulation experiment: spatial resolution is robustly preserved by iterative reconstruction (10% MTF: 2.8-3.0 lp/mm), left-ventricle, cardiac functional metrics can be measured from iodine map segmentations with ~1% error, and small calcifications (615 μm) can be detected during slow moving phases of the cardiac cycle. Given these preliminary results, we believe that PCD technology will enhance dynamic CT imaging applications with high-fidelity spectral and material information.

dc.identifier

PONE-D-19-14774

dc.identifier.issn

1932-6203

dc.identifier.issn

1932-6203

dc.identifier.uri

https://hdl.handle.net/10161/21142

dc.language

eng

dc.publisher

Public Library of Science (PLoS)

dc.relation.ispartof

PloS one

dc.relation.isversionof

10.1371/journal.pone.0218417

dc.subject

Animals

dc.subject

Mice, Knockout

dc.subject

Mice

dc.subject

Image Interpretation, Computer-Assisted

dc.subject

Tomography, X-Ray Computed

dc.subject

Heart Function Tests

dc.subject

Phantoms, Imaging

dc.subject

Photons

dc.subject

Female

dc.subject

X-Ray Microtomography

dc.subject

Four-Dimensional Computed Tomography

dc.title

Photon-counting cine-cardiac CT in the mouse.

dc.type

Journal article

duke.contributor.orcid

Lee, Chang-Lung|0000-0002-0673-633X

duke.contributor.orcid

Badea, Cristian T|0000-0002-1850-2522

pubs.begin-page

e0218417

pubs.issue

9

pubs.organisational-group

School of Medicine

pubs.organisational-group

Biomedical Engineering

pubs.organisational-group

Duke Cancer Institute

pubs.organisational-group

Radiology

pubs.organisational-group

Duke

pubs.organisational-group

Pratt School of Engineering

pubs.organisational-group

Institutes and Centers

pubs.organisational-group

Clinical Science Departments

pubs.publication-status

Published

pubs.volume

14

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Photon-counting cine-cardiac CT in the mouse.pdf
Size:
4.15 MB
Format:
Adobe Portable Document Format
Description:
Accepted version