Bipolar tissue sealant device decreases hemoglobin loss in multilevel spine surgery.

dc.contributor.author

Hill, Steven E

dc.contributor.author

Broomer, Bob

dc.contributor.author

Stover, John

dc.contributor.author

White, William

dc.contributor.author

Richardson, William

dc.date.accessioned

2024-08-15T13:17:50Z

dc.date.available

2024-08-15T13:17:50Z

dc.date.issued

2012-12

dc.description.abstract

Background

Traditional techniques for obtaining hemostasis during orthopedic surgery, such as conventional electrocautery and sealants, have limited clinical effectiveness in reducing hemoglobin (Hb) loss and requirement for transfusion. The bipolar tissue sealant device studied in this trial combines radiofrequency energy with saline irrigation to hemostatically seal both cut bone and soft tissue, potentially aiding hemostasis.

Study design and methods

Sixty patients undergoing multilevel posterior lumbar instrumentation and fusion were randomly assigned to unipolar cautery alone (control group) or unipolar cautery plus use of the bipolar tissue sealant device (treatment group). Hb loss from the surgical field was measured (rather than estimated) and compared between the two groups. The primary hypothesis was that the treatment group would lose significantly less Hb than the control group.

Results

The control group experienced a mean Hb loss of 102.4 g while the treatment group showed a significantly lower mean Hb loss of 66.2 g (p = 0.0004). No significant difference was found between groups with respect to secondary endpoints including length of surgery, number of red blood cell units transfused, number of total blood component units transfused, transfusion avoidance, length of stay, or serious adverse events.

Conclusion

Use of a bipolar tissue sealant device in addition to unipolar cautery significantly decreased Hb loss during multilevel, posterior lumbar spinal instrumentation and fusion when compared with unipolar cautery alone.
dc.identifier.issn

0041-1132

dc.identifier.issn

1537-2995

dc.identifier.uri

https://hdl.handle.net/10161/31395

dc.language

eng

dc.publisher

Wiley

dc.relation.ispartof

Transfusion

dc.relation.isversionof

10.1111/j.1537-2995.2012.03649.x

dc.rights.uri

https://creativecommons.org/licenses/by-nc/4.0

dc.subject

Bone and Bones

dc.subject

Lumbar Vertebrae

dc.subject

Humans

dc.subject

Blood Loss, Surgical

dc.subject

Sodium Chloride

dc.subject

Hemoglobins

dc.subject

Catheter Ablation

dc.subject

Cautery

dc.subject

Hemostasis, Surgical

dc.subject

Spinal Fusion

dc.subject

Prospective Studies

dc.subject

Single-Blind Method

dc.subject

Adult

dc.subject

Aged

dc.subject

Aged, 80 and over

dc.subject

Middle Aged

dc.subject

Female

dc.subject

Male

dc.title

Bipolar tissue sealant device decreases hemoglobin loss in multilevel spine surgery.

dc.type

Journal article

duke.contributor.orcid

Richardson, William|0000-0001-9608-199X|0000-0002-8750-7263|0009-0003-7526-7797

pubs.begin-page

2594

pubs.end-page

2599

pubs.issue

12

pubs.organisational-group

Duke

pubs.organisational-group

School of Medicine

pubs.organisational-group

Clinical Science Departments

pubs.organisational-group

Anesthesiology

pubs.organisational-group

Anesthesiology, Cardiothoracic

pubs.organisational-group

Orthopaedic Surgery

pubs.organisational-group

University Institutes and Centers

pubs.organisational-group

Duke Global Health Institute

pubs.organisational-group

Neurosurgery

pubs.publication-status

Published

pubs.volume

52

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Bipolar tissue sealant device decreases...pdf
Size:
2.51 MB
Format:
Adobe Portable Document Format