Dose coefficients for organ dosimetry in tomosynthesis imaging of adults and pediatrics across diverse protocols.

dc.contributor.author

Sharma, Shobhit

dc.contributor.author

Kapadia, Anuj

dc.contributor.author

Ria, Francesco

dc.contributor.author

Segars, W Paul

dc.contributor.author

Samei, Ehsan

dc.date.accessioned

2022-06-15T16:48:09Z

dc.date.available

2022-06-15T16:48:09Z

dc.date.issued

2022-06-11

dc.date.updated

2022-06-15T16:48:08Z

dc.description.abstract

Purpose

The gold-standard method for estimation of patient-specific organ doses in digital tomosynthesis (DT) requires protocol-specific Monte Carlo (MC) simulations of radiation transport in anatomically accurate computational phantoms. Although accurate, MC simulations are computationally expensive, leading to a turnaround time in the order of core hours for simulating a single exam. This limits their clinical utility. The purpose of this study is to overcome this limitation by utilizing patient- and protocol-specific MC simulations to develop a comprehensive database of air-kerma-normalized organ dose coefficients for a virtual population of adult and pediatric patient models over an expanded set of exam protocols in DT for retrospective and prospective estimation of radiation dose in clinical tomosynthesis.

Materials and methods

A clinically representative virtual population of 14 patient models was used, with pediatric models (M and F) at ages 1, 5, 10, and 15 and adult patient models (M and F) with BMIs at 10th , 50th , and 90th percentiles of the US population. A GPU-based MC simulation framework was used to simulate organ doses in the patient models, incorporating the scanner-specific configuration of a clinical DT system (VolumeRad, GE Healthcare, Waukesha, WI) and an expanded set of exam protocols including 21 distinct acquisition techniques for imaging a variety of anatomical regions (head and neck, thorax, spine, abdomen, and knee). Organ dose coefficients (hn ) were estimated by normalizing organ dose estimates to air kerma at 70 cm (X70cm ) from the source in the scout view. The corresponding coefficients for projection radiography were approximated using organ doses estimated for the scout view. The organ dose coefficients were further used to compute air-kerma-normalized patient-specific effective dose coefficients (Kn ) for all combinations of patients and protocols, and a comparative analysis examining the variation of radiation burden across sex, age, and exam protocols in DT, and with projection radiography was performed.

Results

The database of organ dose coefficients (hn ) containing 294 distinct combinations of patients and exam protocols was developed and made publicly available. The values of Kn were observed to produce estimates of effective dose in agreement with prior studies and consistent with magnitudes expected for pediatric and adult patients across the different exam protocols, with head and neck regions exhibiting relatively lower and thorax and C-spine (apsc, apcs) regions relatively higher magnitudes. The ratios (r = Kn /Kn,rad ) quantifying the differences air-kerma-normalized patient-specific effective doses between DT and projection radiography were centered around 1.0 for all exam protocols, with the exception of protocols covering the knee region (pawk, patk).

Conclusions

This study developed a database of organ dose coefficients for a virtual population of 14 adult and pediatric XCAT patient models over a set of 21 exam protocols in DT. Using empirical measurements of air kerma in the clinic, these organ dose coefficients enable practical retrospective and prospective patient-specific radiation dosimetry. The computation of air-kerma-normalized patient-specific effective doses further enable the comparison of radiation burden to the patient populations between protocols and between imaging modalities (e.g., DT and projection radiography), as presented in this study. This article is protected by copyright. All rights reserved.
dc.identifier.issn

0094-2405

dc.identifier.issn

2473-4209

dc.identifier.uri

https://hdl.handle.net/10161/25118

dc.language

eng

dc.publisher

Wiley

dc.relation.ispartof

Medical physics

dc.relation.isversionof

10.1002/mp.15798

dc.subject

Task Group 321

dc.subject

XCAT

dc.subject

monte carlo

dc.subject

patient specific

dc.subject

radiation dosimetry & risk

dc.subject

tomosynthesis

dc.title

Dose coefficients for organ dosimetry in tomosynthesis imaging of adults and pediatrics across diverse protocols.

dc.type

Journal article

duke.contributor.orcid

Ria, Francesco|0000-0001-5902-7396

duke.contributor.orcid

Samei, Ehsan|0000-0001-7451-3309

pubs.organisational-group

Duke

pubs.organisational-group

School of Medicine

pubs.organisational-group

Staff

pubs.organisational-group

Clinical Science Departments

pubs.organisational-group

Radiology

pubs.publication-status

Published

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Tomosynthesis_organ_dosimetry_Sharma_et_al_2022_Medical_Physics.pdf
Size:
2.74 MB
Format:
Adobe Portable Document Format
Description:
Accepted version