Evaluation of Quantitative Potential of Breast Tomosynthesis Using a Voxelized Anthropomorphic Breast Phantom

dc.contributor.advisor

Lo, Joseph YuanChieh

dc.contributor.author

Mehtaji, Deep Sunil

dc.date.accessioned

2010-05-13T17:50:13Z

dc.date.available

2011-07-26T04:30:04Z

dc.date.issued

2010

dc.department

Medical Physics

dc.description.abstract

Purpose: To assess the quantitative potential of breast tomosynthesis by estimating the percent density of voxelized anthropomorphic breast phantoms.

Methods and Materials:A Siemens breast tomosynthesis system was modeled using Monte Carlo methods and a voxelized anthropomorphic breast phantom. The images generated by the simulation were reconstructed using Siemens filtered back-projection software. The non-uniform background due to scatter, heel effect, and limited angular sampling was estimated by simulating and subtracting images of a uniform 100% fatty breast phantom. To estimate the density of each slice, the total number of fatty and glandular voxels was calculated both before and after applying a thresholding algorithm to classify voxels as fat vs. glandular. Finally, the estimated density of the reconstructed slice was compared to the known percent density of the corresponding slice from the voxelized phantom. This percent density estimation comparison was done for a 35%- and a 60%-dense 5cm breast phantom.

Results: Without thresholding, overall density estimation errors for the central eleven slices were 4.97% and 2.55% for the 35% and 60% dense phantoms, respectively. After thresholding to classify voxels as fat vs. glandular, errors for central eleven were 7.99% and 6.26%, respectively. Voxel to voxel matching of the phantom vs. reconstructed slice demonstrated 75.69% and 75.25% respectively of voxels were correctly classified.

Conclusion: The errors in slice density estimation were <8% for both the phantoms thus implying that quantification of breast density using tomosynthesis is possible. However, limitations of the acquisition and reconstruction process continue to pose challenges in density estimation leading to potential voxel to voxel errors that warrant further investigation.

dc.identifier.uri

https://hdl.handle.net/10161/2502

dc.language.iso

en_US

dc.subject

Physics, Fluid and Plasma

dc.subject

Biomedical engineering

dc.subject

antropomorphic

dc.subject

Breast imaging

dc.subject

density estimation

dc.subject

Phantom

dc.subject

quantitation

dc.subject

Tomosynthesis

dc.title

Evaluation of Quantitative Potential of Breast Tomosynthesis Using a Voxelized Anthropomorphic Breast Phantom

dc.type

Master's thesis