PERFORMANCE EVALUATION OF THE NEW OTTO BOCK “DynamicArm” BY MEANS OF BIOMECHANICAL MODELLING

dc.contributor.author

Cutti, Andrea G.

dc.contributor.author

Davalli, Angelo

dc.contributor.author

Valeria, Gazzotti

dc.contributor.author

Andrei, Ninu

dc.date.accessioned

2010-07-20T19:22:01Z

dc.date.available

2010-07-20T19:22:01Z

dc.date.issued

2005

dc.description.abstract

Through infra-red motion analysis systems it is possible to acquire the 3D joint kinematics of a patient while performing every day activities. These data, combined with a biomechanical model of the anatomical structures under investigation and clinical rating scales, can form the basis for an objective assessment of the patient motor ability. When the subject acquired is an amputee fitted with a new prosthetic arm, the information provided can be useful not only for the practitioner but also for the prosthesis designer. The aim of this work is to give an example of this kind of clinical/technology assessment, presenting the results obtained for a young trans-humeral amputee fitted with a prototype of the new Otto Bock DynamicArm. In particular, the analysis intended to quantitatively evaluate: 1) the performances of the Otto Bock arm, and in particular of the electromechanic elbow, when controlled in-vivo by the patient EMG signals; 2) how the patient controls the prosthesis, in order to identify critical movements and prevent possible disorders; 3) if the new prosthesis increases the patient abilities.

dc.format.extent

184267 bytes

dc.format.mimetype

application/pdf

dc.identifier.citation

Proceedings of the MEC’05 conference, UNB; 2005.

dc.identifier.uri

https://hdl.handle.net/10161/2702

dc.language.iso

en_US

dc.publisher

Myoelectric Symposium

dc.subject

Dynamic Arm

dc.subject

Biomechanics

dc.title

PERFORMANCE EVALUATION OF THE NEW OTTO BOCK “DynamicArm” BY MEANS OF BIOMECHANICAL MODELLING

dc.type

Other article

Files