Item Open Access Evaluation of a Dedicated SPECT-CT Mammotomography System for Quantitative Hybrid Breast Imaging(2010) Cutler, Spencer JohnsonThe overall goal of this dissertation is to optimize and evaluate the performance of the single photon emission computed tomography (SPECT) subsystem of a dedicated three-dimensional (3D) dual-modality breast imaging system for enhanced semi-automated, quantitative clinical imaging. This novel hybrid imaging system combines functional or molecular information obtained with a SPECT subsystem with high-resolution anatomical imaging obtained with a low dose x-ray Computed Tomography (CT) subsystem. In this new breast imaging paradigm, coined "mammotomography," the subject is imaged lying prone while the individual subsystems sweep 3-dimensionally about her uncompressed, pendant breast, providing patient comfort compared to traditional compression-based imaging modalities along with high fidelity and information rich images for the clinician.
System evaluation includes a direct comparison between dedicated 3D SPECT and dedicated 2D scintimammography imaging using the same high performance, semi-conductor gamma camera. Due to the greater positioning flexibility of the SPECT system gantry, under a wide range of measurement conditions, statistically significantly (p<0.05) more lesions and smaller lesion sizes were detected with dedicated breast SPECT than with compressed breast scintimammography. The importance of good energy resolution for uncompressed SPECT breast imaging was also investigated. Results clearly illustrate both visual and quantitative differences between the various energy windows, with energy windows slightly wider than the system resolution having the best image contrast and quality.
An observer-based contrast-detail study was performed in an effort to evaluate the limits of object detectability under various imaging conditions. The smallest object detail was observed using a 45-degree tilted trajectory acquisition. The complex 3D projected sine wave acquisition, however, had the most consistent combined intra and inter-observer results, making it potentially the best imaging approach for consistent clinical imaging.
Automatic ROR contouring is implemented using a dual-layer light curtain design, ensuring that an arbitrarily shaped breast is within ~1 cm of the camera face, but no closer than 0.5 cm at every projection angle of a scan. Autocontouring enables simplified routine scanning using complex 3D trajectories, and yields improved image quality. Absolute quantification capabilities are also integrated into the SPECT system, allowing the calculation of in vivo total lesion activity. Initial feasibility studies in controlled low noise experiments show promising results with total activity agreement within 10% of the dose calibrator values.
The SPECT system is integrated with a CT scanner for added diagnostic power. Initial human subject studies demonstrate the clinical potential of the hybrid SPECT-CT breast imaging system. The reconstructed SPECT-CT images illustrate the power of fusing functional SPECT information to localize lesions not easily seen in the anatomical CT images. Enhanced quantitative 3D SPECT-CT breast imaging, now with the ability to dynamically contour any sized breast, has high potential to improve detection, diagnosis, and characterization of breast cancer in upcoming larger-scale clinical testing.
Item Open Access Investigating Functional Breast Image Quality and Quantification with a Dedicated SPECT-CT System(2011) Perez, Kristy LynnThis work investigates phantom and subject positioning as well as collecting data with a variety of angular sampling and acquisition trajectories. The overall goal of this work has been to utilize the dedicated, breast SPECT-CT system to acquire the best possible images. A large portion of this work has been to apply corrections to the system for quantitative imaging. The system has been shown to provide high quality images with minimal out-of-field signal contribution. Additionally, the quantification procedure has been shown to be within 10% of the known activity concentration present at the time of imaging for both VAOR and PROJSINE trajectories.
Item Open Access Investigation and Development of a Fully 3D Tilt Capable Hybrid SPECT - CT System for Dedicated Breast Imaging(2015) Shah, JainilX-ray mammography has been the gold standard for breast imaging for decades, despite the significant limitations posed by the two dimensional (2D) image acquisitions. Difficulty in diagnosing lesions close to the chest wall and axilla, high amount of structural overlap and patient discomfort due to compression are only some of these limitations. To overcome these drawbacks, three dimensional (3D) breast imaging modalities have been developed including dual modality single photon emission computed tomography (SPECT) and computed tomography (CT) systems. This thesis focuses on the development and integration of the next generation of such a device for dedicated breast imaging. The goals of this dissertation work are to: [1] understand and characterize any effects of fully 3-D trajectories on reconstructed image scatter correction, absorbed dose and Hounsifeld Unit accuracy, and [2] design, develop and implement the fully flexible, third generation hybrid SPECT-CT system capable of traversing complex 3D orbits about a pendant breast volume, without interference from the other. Such a system would overcome artifacts resulting from incompletely sampled divergent cone beam imaging schemes and allow imaging closer to the chest wall, which other systems currently under research and development elsewhere cannot achieve.
The dependence of x-ray scatter radiation on object shape, size, material composition and the CT acquisition trajectory, was investigated with a well-established beam stop array (BSA) scatter correction method. While the 2D scatter to primary ratio (SPR) was the main metric used to characterize total system scatter, a new metric called ‘normalized scatter contribution’ was developed to compare the results of scatter correction on 3D reconstructed volumes. Scatter estimation studies were undertaken with a sinusoidal saddle (±15° polar tilt) orbit and a traditional circular (AZOR) orbit. Clinical studies to acquire data for scatter correction were used to evaluate the 2D SPR on a small set of patients scanned with the AZOR orbit. Clinical SPR results showed clear dependence of scatter on breast composition and glandular tissue distribution, otherwise consistent with the overall phantom-based size and density measurements. Additionally, SPR dependence was also observed on the acquisition trajectory where 2D scatter increased with an increase in the polar tilt angle of the system.
The dose delivered by any imaging system is of primary importance from the patient’s point of view, and therefore trajectory related differences in the dose distribution in a target volume were evaluated. Monte Carlo simulations as well as physical measurements using radiochromic film were undertaken using saddle and AZOR orbits. Results illustrated that both orbits deliver comparable dose to the target volume, and only slightly differ in distribution within the volume. Simulations and measurements showed similar results, and all measured dose values were within the standard screening mammography-specific, 6 mGy dose limit, which is used as a benchmark for dose comparisons.
Hounsfield Units (HU) are used clinically in differentiating tissue types in a reconstructed CT image, and therefore the HU accuracy of a system is very important, especially when using non-traditional trajectories. Uniform phantoms filled with various uniform density fluids were used to investigate differences in HU accuracy between saddle and AZOR orbits. Results illustrate the considerably better performance of the saddle orbit, especially close to the chest and nipple region of what would clinically be a pedant breast volume. The AZOR orbit causes shading artifacts near the nipple, due to insufficient sampling, rendering a major portion of the scanned phantom unusable, whereas the saddle orbit performs exceptionally well and provides a tighter distribution of HU values in reconstructed volumes.
Finally, the third generation, fully-suspended SPECT-CT system was designed in and developed in our lab. A novel mechanical method using a linear motor was developed for tilting the CT system. A new x-ray source and a custom made 40 x 30 cm2 detector were integrated on to this system. The SPECT system was nested, in the center of the gantry, orthogonal to the CT source-detector pair. The SPECT system tilts on a goniometer, and the newly developed CT tilting mechanism allows ±15° maximum polar tilting of the CT system. The entire gantry is mounted on a rotation stage, allowing complex arbitrary trajectories for each system, without interference from the other, while having a common field of view. This hybrid system shows potential to be used clinically as a diagnostic tool for dedicated breast imaging.