Electrochemical Burns: A Risk With Myoelectric Prostheses
Abstract
The authors describe their experience myoelectric prostheses, with incidents of minor trauma, to users of which is thought to be associated with electrochemical bums arising from a dc potential existing between electrodes. These incidents have occurred over the period 1968 - 1996, in Winnipeg, Toronto and Fredericton. A possible mechanism is suggested for electrochemical burns where a dc potential of one volt or more exists. Finally, recommendations are made for increased vigilance concerning this possible problem.
Type
Department
Description
Provenance
Subjects
Citation
From "MEC 97," Proceedings of the 1997 MyoElectric Controls/Powered Prosthetics Symposium Fredericton, New Brunswick, Canada: August, 1997. Copyright University of New Brunswick.
Permalink
Collections
Copyright 2002, 2005 and 2008, The University of New Brunswick.