CONE BEAM COMPUTED TOMOGRAPHY (CBCT) DOSIMETRY: MEASUREMENTS AND MONTE CARLO SIMULATIONS

Loading...
Thumbnail Image

Date

2010

Journal Title

Journal ISSN

Volume Title

Repository Usage Stats

1404
views
8144
downloads

Abstract

Cone beam computed tomography (CBCT) is a 3D x-ray imaging technique in which the x-ray beam is transmitted to an object with wide beam geometry producing a 2D image per projection. Due to its faster image acquisition time, wide coverage length per scan, and fewer motion artifacts, the CBCT system is rapidly replacing the conventional CT system and becoming popular in diagnostic and therapeutic radiology. However, there are few studies performed in CBCT dosimetry because of the absence of a standard dosimetric protocol for CBCT. Computed tomography dose index (CTDI), a standardized metric in conventional CT dosimetry, or direct organ dose measurements have been limitedly used in the CBCT dosimetry.

This dissertation investigated the CBCT dosimetry from the CTDI method to the organ, effective dose, risk estimations with physical measurements and Monte Carlo (MC) simulations.

An On-Board Imager (OBI, Varian Medical Systems, Palo Alto, CA) was used to perform old and new CBCT scan protocols. The new CBCT protocols introduced both partial and full angle scan modes while the old CBCT protocols only used the full angle mode. A metal-oxide-semiconductor-field-effect transistor (MOSFET) and an ion chamber were employed to measure the cone beam CTDI (CTDICB) in CT phantoms and organ dose in a 5-year-old pediatric anthropomorphic phantom. Radiochromic film was also employed to measure the axial dose profiles. A point dose method was used in the CTDI estimation.

The BEAMnrc/EGSnrc MC system was used to simulate the CBCT scans; the MC model of the OBI x-ray tube was built into the system and validated by measurements characterizing the cone beam quality in the aspects of the x-ray spectrum, half value layer (HVL) and dose profiles for both full-fan and half-fan modes. Using the validated MC model, CTDICB, dose profile integral (DPI), cone beam dose length product (DLPCB), and organ doses were calculated with voxelized MC CT phantoms or anthropomorphic phantoms. Effective dose and radiation risks were estimated from the organ dose results.

The CTDICB of the old protocols were found to be 84 and 45 mGy for standard dose, head and body protocols. The CTDICB of the new protocols were found to be 6.0, 3.2, 29.0, 25.4, 23.8, and 7.7 mGy for the standard dose head, low dose head, high quality head, pelvis, pelvis spotlight, and low dose thorax protocols respectively. The new scan protocols were found to be advantageous in reducing the patient dose while offering acceptable image quality.

The mean effective dose (ED) was found to be 37.8 ±0.7 mSv for the standard head and 8.1±0.2 mSv for the low dose head protocols (old) in the 5-year-old phantom. The lifetime attributable risk (LAR) of cancer incidence ranged from 23 to 144 cases per 100,000 exposed persons for the standard-dose mode and from five to 31 cases per 100,000 exposed persons for the low-dose mode. The relative risk (RR) of cancer incidence ranged from 1.003 to 1.054 for the standard-dose mode and from 1.001 to 1.012 for the low-dose mode.

The MC method successfully estimated the CTDICB, organ and effective dose despite the heavy calculation time. The point dose method was found to be capable of estimating the CBCT dose with reasonable accuracy in the clinical environment.

Description

Provenance

Citation

Citation

Kim, Sangroh (2010). CONE BEAM COMPUTED TOMOGRAPHY (CBCT) DOSIMETRY: MEASUREMENTS AND MONTE CARLO SIMULATIONS. Dissertation, Duke University. Retrieved from https://hdl.handle.net/10161/2371.

Collections


Except where otherwise noted, student scholarship that was shared on DukeSpace after 2009 is made available to the public under a Creative Commons Attribution / Non-commercial / No derivatives (CC-BY-NC-ND) license. All rights in student work shared on DukeSpace before 2009 remain with the author and/or their designee, whose permission may be required for reuse.