Evaluation of Centrally Located Sources in Coincidence Timing Calibration for Time-of-Flight PET

dc.contributor.advisor

Turkington, Timothy G

dc.contributor.author

Wargo, Richard Ryan

dc.date.accessioned

2012-05-25T20:13:07Z

dc.date.available

2012-05-25T20:13:07Z

dc.date.issued

2012

dc.department

Medical Physics

dc.description.abstract

Coincidence Timing Calibration (CTC) is an essential part of ensuring proper PET scanner function. The purpose of CTC is to account for timing differences in detector modules. The importance and precision in which this calibration needs to work is even more stringent for Time-of-Flight (TOF) PET. In this work, we looked to investigate the CTC process by which the TOF capable GE PET/CT Discovery-690 (D690) operates. Currently, it uses a 68Ge rotating pin source (RPS) to perform the calibration. The purpose of this work was to investigate the use of a centrally located source to perform the calibration. The timing resolution of the D690 was determined and used as a metric to evaluate both methods.

Two cylindrical 18F filled phantoms of 7.5 and 10 cm diameter were used to perform the CTC. The RPS and system table motion had to be disabled in order to use the centrally located sources in the CTC. All CTCs started with the default calibration file in place. Iterations of the CTC were performed until convergence of the calibration was observed on the review screen. Even after convergence, more iterations were performed for further analysis. At the end of the CTC with the centrally located sources, a follow-up iteration with the RPS was performed to see what adjustments would be made. Next, the timing resolution of the system was measured using a 68Ge line source. An apparatus with known locations to support the source allowed for the evaluation of the timing resolution off the central axis. The importance of this was that it allowed for non-centrally located lines of response to be evaluated. Furthermore, the timing resolution was measured with specific calibration files enabled that corresponded to particular iterations. In addition, a novel way of measuring the timing resolution (propagated method) for a particular calibration result without an actual measurement with that calibration enabled was developed and implemented. This greatly reduced the number of resolution measurements needed, which was particularly helpful for evaluating the improvement for each iteration.

The timing resolution of the system improved as more iterations were done. The difference between the propagated and measured timing resolution was under 2% most of the time. The cases in which the discrepancy was larger than 2% corresponded to one of the first iterations performed. After 15 iterations were performed for both centrally located scanners, the timing resolution of the system was measured through propagation to be 610 ps. The 15 iterations amounts to 15 minutes of acquisition time. After one iteration for the RPS, the timing resolution was measured to be 585 ps (587 ps in the propagated measurement). The single iteration of the RPS corresponded to 8 minutes of acquisition time. When following up the final iterations of the centrally located sources with the RPS, there was a change observed that improved the timing resolution to that measured after only one iteration of the RPS.

Conclusively, trends in the data showed that the centrally located sources did bring opposing detectors into good timing alignment with one another. These trends also indicated that the current CTC algorithm is not optimized for centrally located sources for the diameters tested. Finally, the method of propagating the change in calibration files illustrated a new CTC process method.

dc.identifier.uri

https://hdl.handle.net/10161/5479

dc.subject

Medical imaging and radiology

dc.subject

Coincidence Timing Calibration

dc.subject

PET

dc.subject

Time-of-Fight

dc.subject

Timing Resolution

dc.title

Evaluation of Centrally Located Sources in Coincidence Timing Calibration for Time-of-Flight PET

dc.type

Master's thesis