Label-free Biodetection with Individual Plasmonic Nanoparticles

dc.contributor.advisor

Chilkoti, Ashutosh

dc.contributor.author

Nusz, Gregory

dc.date.accessioned

2010-05-10T19:53:01Z

dc.date.available

2011-07-26T04:30:04Z

dc.date.issued

2010

dc.department

Biomedical Engineering

dc.description.abstract

The refractive index sensitivity of plasmonic nanoparticles is utilized in the development of real-time, label-free biodetection. Analyte molecules that bind to receptor-conjugated nanoparticles cause an increase in local refractive index that in turn induces an energy shift in the optical resonance of the particle. Biomolecular binding is quantified by quantitatively measuring these resonance shifts. This work describes the application and optimization of a biomolecular detection system based on gold nanorods as an optical transducer.

A microspectroscopy system was developed to collect scattering spectra of single nanoparticles, and measure shifts of the spectra as a function of biomolecular binding. The measurement uncertainty of LSPR peak shifts of the system was demonstrated to be 0.3 nm. An analytical model was also developed that provides the optimal gold nanorod geometry for detection with specified receptor-analyte pair. The model was applied to the model biotin-streptavidin system, which resulted in sensing system with a detection limit of 130 pM - an improvement by four orders of magnitude over any other single-particle biodetection previously presented in the literature.

Alternative optical detection schemes were also investigated that could facilitate mulitplexed biosensing. A theoretical model was built to investigate the efficacy of using a multi-channel detector analogous to a conventional RGB camera. The results of the model indicated that even in the best case, the detection capabilities of such a system did not provide advantages over the microspectroscopic approach.

We presented a novel hyperspectral detection scheme we term Dual-Order Spectral Imaging (DOSI) which is capable of simultaneously measuring spectra of up to 160 individual regions within a microscope's field of view. This technique was applied to measuring shifts of individual nanoparticles and was found to have a peak measurement uncertainty of 1.29 nm, at a measurement rate of 2-5 Hz.

dc.identifier.uri

https://hdl.handle.net/10161/2264

dc.language.iso

en_US

dc.subject

Biomedical engineering

dc.subject

Physics, Optics

dc.subject

Biodetection

dc.subject

Biosensor

dc.subject

Label-free

dc.subject

Nanoparticle

dc.subject

Plasmon

dc.title

Label-free Biodetection with Individual Plasmonic Nanoparticles

dc.type

Dissertation

duke.embargo.months

12

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
D_Nusz_Gregory_a_201005.pdf
Size:
13.77 MB
Format:
Adobe Portable Document Format

Collections