An automated method for comparing motion artifacts in cine four-dimensional computed tomography images.

Loading...
Thumbnail Image

Date

2012-11-08

Journal Title

Journal ISSN

Volume Title

Repository Usage Stats

66
views
29
downloads

Citation Stats

Abstract

The aim of this study is to develop an automated method to objectively compare motion artifacts in two four-dimensional computed tomography (4D CT) image sets, and identify the one that would appear to human observers with fewer or smaller artifacts. Our proposed method is based on the difference of the normalized correlation coefficients between edge slices at couch transitions, which we hypothesize may be a suitable metric to identify motion artifacts. We evaluated our method using ten pairs of 4D CT image sets that showed subtle differences in artifacts between images in a pair, which were identifiable by human observers. One set of 4D CT images was sorted using breathing traces in which our clinically implemented 4D CT sorting software miscalculated the respiratory phase, which expectedly led to artifacts in the images. The other set of images consisted of the same images; however, these were sorted using the same breathing traces but with corrected phases. Next we calculated the normalized correlation coefficients between edge slices at all couch transitions for all respiratory phases in both image sets to evaluate for motion artifacts. For nine image set pairs, our method identified the 4D CT sets sorted using the breathing traces with the corrected respiratory phase to result in images with fewer or smaller artifacts, whereas for one image pair, no difference was noted. Two observers independently assessed the accuracy of our method. Both observers identified 9 image sets that were sorted using the breathing traces with corrected respiratory phase as having fewer or smaller artifacts. In summary, using the 4D CT data of ten pairs of 4D CT image sets, we have demonstrated proof of principle that our method is able to replicate the results of two human observers in identifying the image set with fewer or smaller artifacts.

Department

Description

Provenance

Citation

Published Version (Please cite this version)

10.1120/jacmp.v13i6.3838

Publication Info

Cui, Guoqiang, Brian Jew, Julian C Hong, Eric W Johnston, Billy W Loo and Peter G Maxim (2012). An automated method for comparing motion artifacts in cine four-dimensional computed tomography images. Journal of applied clinical medical physics, 13(6). p. 3838. 10.1120/jacmp.v13i6.3838 Retrieved from https://hdl.handle.net/10161/19086.

This is constructed from limited available data and may be imprecise. To cite this article, please review & use the official citation provided by the journal.

Scholars@Duke

Cui

Guoqiang Cui

Assistant Professor of Radiation Oncology

Clinical Research in automated treatment planning in Radiosurgery and SBRT, Image Guided Radiation Therapy (IGRT), and Quality Assurance (QA) in Radiation Therapy.


Unless otherwise indicated, scholarly articles published by Duke faculty members are made available here with a CC-BY-NC (Creative Commons Attribution Non-Commercial) license, as enabled by the Duke Open Access Policy. If you wish to use the materials in ways not already permitted under CC-BY-NC, please consult the copyright owner. Other materials are made available here through the author’s grant of a non-exclusive license to make their work openly accessible.