Removal of ECG Artifacts from Myoelectric Prosthesis Control Signals

dc.contributor.author

Zhou, Ping

dc.contributor.author

Lowery, Madeleine M.

dc.contributor.author

Weir, Richard F.

dc.contributor.author

Kuiken, Todd A.

dc.date.accessioned

2010-07-22T19:39:37Z

dc.date.available

2010-07-22T19:39:37Z

dc.date.issued

2005

dc.description.abstract

We investigated removal of electrocardiogram (ECG) artifacts from the myoelectric prosthesis control signals, taken from the reinnervated pectoralis muscles of a patient with bilateral amputations at shoulder disarticulation level. The performance of various ECG artifact removal methods including high pass filtering, spike clipping, template subtracting, wavelet thresholding and adaptive filtering was presented. In particular, considering the clinical requirements and memory limitation of commercial prosthesis controllers, we further explored suitable means of ECG artifact removal for clinical application.

dc.format.extent

115310 bytes

dc.format.mimetype

application/pdf

dc.identifier.citation

Proceedings of the MEC’05 conference, UNB; 2005.

dc.identifier.uri

https://hdl.handle.net/10161/2759

dc.language.iso

en_US

dc.publisher

Myoelectric Symposium

dc.subject

myoelectric signals

dc.subject

myoelectric prosthesis

dc.title

Removal of ECG Artifacts from Myoelectric Prosthesis Control Signals

dc.type

Other article