Target localization using scanner-acquired SPECT data.

dc.contributor.author

Roper, Justin R

dc.contributor.author

Bowsher, James E

dc.contributor.author

Wilson, Joshua M

dc.contributor.author

Turkington, Timothy G

dc.contributor.author

Yin, Fang-Fang

dc.date.accessioned

2019-10-04T16:16:59Z

dc.date.available

2019-10-04T16:16:59Z

dc.date.issued

2012-05-10

dc.date.updated

2019-10-04T16:16:57Z

dc.description.abstract

Target localization using single photon emission computed tomography (SPECT) and planar imaging is being investigated for guiding radiation therapy delivery. Previous studies on SPECT-based localization have used computer-simulated or hybrid images with simulated tumors embedded in disease-free patient images where the tumor position is known and localization can be calculated directly. In the current study, localization was studied using scanner-acquired images. Five fillable spheres were placed in a whole body phantom. Sphere-to-background 99mTc radioactivity was 6:1. Ten independent SPECT scans were acquired with a Trionix Triad scanner using three detector trajectories: left lateral 180°, 360°, and right lateral 180°. Scan time was equivalent to 4.5 min. Images were reconstructed with and without attenuation correction. True target locations were estimated from 12 hr SPECT and CT images. From the 12 hr SPECT scan, 45 sets of orthogonal planar images were used to assess target localization; total acquisition time per set was equivalent to 4.5min. A numerical observer localized the center of the targets in the 4.5 min SPECT and planar images. SPECT-based localization errors were compared for the different detector trajectories. Across the four peripheral spheres, and using optimal iteration numbers and postreconstruction smoothing, means and standard deviations in localization errors were 0.90 ± 0.25 mm for proximal 180° trajectories, 1.31 ± 0.51 mm for 360° orbits, and 3.93 ± 1.48 mm for distal 180° trajectories. This rank order in localization performance is predicted by target attenuation and distance from the target to the collimator. For the targets with mean localization errors < 2 mm, attenuation correction reduced localization errors by 0.15 mm on average. The improvement from attenuation correction was 1.0 mm on average for the more poorly localized targets. Attenuation correction typically reduced localization errors, but for well-localized targets, the detector trajectory generally had a larger effect. Localization performance was found to be robust to iteration number and smoothing. Localization was generally worse using planar images as compared with proximal 180° and 360° SPECT scans. Using a proximal detector trajectory and attenuation correction, localization errors were within 2 mm for the three superficial targets, thus supporting the current role in biopsy and surgery, and demonstrating the potential for SPECT imaging inside radiation therapy treatment rooms.

dc.identifier.issn

1526-9914

dc.identifier.issn

1526-9914

dc.identifier.uri

https://hdl.handle.net/10161/19405

dc.language

eng

dc.publisher

MULTIMED INC

dc.relation.ispartof

Journal of applied clinical medical physics

dc.relation.isversionof

10.1120/jacmp.v13i3.3724

dc.subject

Humans

dc.subject

Tomography, Emission-Computed, Single-Photon

dc.subject

Image Enhancement

dc.subject

Phantoms, Imaging

dc.subject

Algorithms

dc.subject

Radiotherapy, Image-Guided

dc.title

Target localization using scanner-acquired SPECT data.

dc.type

Journal article

duke.contributor.orcid

Wilson, Joshua M|0000-0002-4175-6301

duke.contributor.orcid

Yin, Fang-Fang|0000-0002-2025-4740|0000-0003-1064-2149

pubs.begin-page

3724

pubs.issue

3

pubs.organisational-group

School of Medicine

pubs.organisational-group

Duke

pubs.organisational-group

Duke Kunshan University Faculty

pubs.organisational-group

Duke Kunshan University

pubs.organisational-group

Duke Cancer Institute

pubs.organisational-group

Institutes and Centers

pubs.organisational-group

Radiation Oncology

pubs.organisational-group

Clinical Science Departments

pubs.organisational-group

Staff

pubs.organisational-group

Duke University Medical Physics Graduate Program

pubs.organisational-group

Radiology

pubs.publication-status

Published

pubs.volume

13

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Target localization using scanner-acquired SPECT data.pdf
Size:
1.53 MB
Format:
Adobe Portable Document Format