Spine SBRT With Halcyon™: Plan Quality, Modulation Complexity, Delivery Accuracy, and Speed.

dc.contributor.author

Petroccia, Heather M

dc.contributor.author

Malajovich, Irina

dc.contributor.author

Barsky, Andrew R

dc.contributor.author

Ghiam, Alireza Fotouhi

dc.contributor.author

Jones, Joshua

dc.contributor.author

Wang, Chunhao

dc.contributor.author

Zou, Wei

dc.contributor.author

Teo, Boon-Keng Kevin

dc.contributor.author

Dong, Lei

dc.contributor.author

Metz, James M

dc.contributor.author

Li, Taoran

dc.date.accessioned

2020-03-24T13:17:45Z

dc.date.available

2020-03-24T13:17:45Z

dc.date.issued

2019-01

dc.date.updated

2020-03-24T13:17:43Z

dc.description.abstract

Purpose: Spine SBRT requires treatment plans with steep dose gradients and tight limits to the cord maximal dose. A new dual-layer staggered 1-cm MLC in Halcyon™ treatment platform has improved leakage, speed, and DLG compared to 120-Millennium (0.5-cm) and High-Definition (0.25-cm) MLCs in the TrueBeam platform. Halcyon™ 2.0 with SX2 MLC modulates fluence with the upper and lower MLCs, while in Halcyon™ 1.0 with SX1 only the lower MLC modulates the fluence and the upper MLC functions as a back-up jaw. We investigated the effects of four MLC designs on plan quality for spine SBRT treatments. Methods: 15 patients previously treated at our institution were re-planned according to the NRG-BR-002 guidelines with a prescription of 3,000 cGy in 3 fractions, 6xFFF, 800 MU/min, and 3-arc VMAT technique. Planning objectives were adjusted manually by an experienced planner to generate optimal plans and kept the same for different MLCs within the same platform. Results: All treatment plans were able to achieve adequate target coverage while meeting NRG-BR002 dosimetric constraints. Planning parameters were evaluated including: conformity index, homogeneity index, gradient measure, and global point dose maximum. Delivery accuracy, modulation complexity, and delivery time were also analyzed for all MLCs. Conclusion: The Halcyon™ dual-layer MLC can generate comparable and clinically equivalent spine SBRT plans to TrueBeam plans with less rapid dose fall-off and lower conformity. MLC width leaf can impact maximum dose to organs at risk and plan quality, but does not cause limitations in achieving acceptable plans for spine SBRT treatments.

dc.identifier.issn

2234-943X

dc.identifier.issn

2234-943X

dc.identifier.uri

https://hdl.handle.net/10161/20267

dc.language

eng

dc.publisher

Frontiers Media SA

dc.relation.ispartof

Frontiers in Oncology

dc.relation.isversionof

10.3389/fonc.2019.00319

dc.subject

Halcyon™

dc.subject

SBRT

dc.subject

modulation complexity score

dc.subject

multi-leaf collimator

dc.subject

plan quality

dc.subject

spinal metastasis

dc.title

Spine SBRT With Halcyon™: Plan Quality, Modulation Complexity, Delivery Accuracy, and Speed.

dc.type

Journal article

pubs.begin-page

319

pubs.issue

APR

pubs.organisational-group

School of Medicine

pubs.organisational-group

Duke Cancer Institute

pubs.organisational-group

Radiation Oncology

pubs.organisational-group

Duke

pubs.organisational-group

Institutes and Centers

pubs.organisational-group

Clinical Science Departments

pubs.publication-status

Published

pubs.volume

9

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Spine SBRT With Halcyon™ Plan Quality, Modulation Complexity, Delivery Accuracy, and Speed.pdf
Size:
3.09 MB
Format:
Adobe Portable Document Format