Laboratory accuracy improvement in the uk neqas leucocyte immunophenotyping immune monitoring program: An eleven-year review via longitudinal mixed effects modeling.

dc.contributor.author

Bainbridge, John

dc.contributor.author

Rountree, Wes

dc.contributor.author

Louzao, Raul

dc.contributor.author

Wong, John

dc.contributor.author

Whitby, Liam

dc.contributor.author

Denny, Thomas N

dc.contributor.author

Barnett, David

dc.coverage.spatial

United States

dc.date.accessioned

2017-06-01T18:30:14Z

dc.date.available

2017-06-01T18:30:14Z

dc.date.issued

2017-05-08

dc.description.abstract

BACKGROUND: The United Kingdom National External Quality Assessment Service (UK NEQAS) for Leucocyte Immunophenotyping Immune Monitoring Programme, provides external quality assessment (EQA) to non-U.S. laboratories affiliated with the NIH NIAID Division of AIDS (DAIDS) clinical trials networks. Selected laboratories are required to have oversight, performance monitoring, and remediation undertaken by Immunology Quality Assessment (IQA) staff under the DAIDS contract. We examined whether laboratory accuracy improves with longer EQA participation and whether IQA remediation is effective. METHODS: Laboratory accuracy, defined by the measurement residuals from trial sample medians, was measured on four outcomes: both CD4+ absolute counts (cells/μL) and percentages; and CD8+ absolute counts (cells/μL) and percentages. Three laboratory categories were defined: IQA monitored (n = 116), United Kingdom/non-DAIDS (n = 137), and non-DAIDS/non-UK (n = 1034). For absolute count outcomes, the groups were subdivided into single platform and dual platform users. RESULTS: Increasing EQA duration was found to be associated with increasing accuracy for all groups in all four lymphocyte subsets (P < 0.0001). In the percentage outcomes, the typical IQA group laboratory improved faster than laboratories from the other two groups (P < 0.005). No difference in the overall rate of improvement was found between groups for absolute count outcomes. However, in the DPT subgroup the IQA group ultimately showed greater homogeneity. CONCLUSIONS: EQA participation coupled with effective laboratory monitoring and remedial action is strongly associated with improved laboratory accuracy, both incrementally and in the proportion of laboratories meeting suggested standards. Improvement in accuracy provides more reliable laboratory information facilitating more appropriate patient treatment decisions. © 2017 International Clinical Cytometry Society.

dc.identifier

https://www.ncbi.nlm.nih.gov/pubmed/28480599

dc.identifier.eissn

1552-4957

dc.identifier.uri

https://hdl.handle.net/10161/14663

dc.language

eng

dc.publisher

Wiley

dc.relation.ispartof

Cytometry B Clin Cytom

dc.relation.isversionof

10.1002/cyto.b.21531

dc.subject

flow cytometry

dc.subject

mixed effects model

dc.subject

quality assessment

dc.subject

quality improvement

dc.title

Laboratory accuracy improvement in the uk neqas leucocyte immunophenotyping immune monitoring program: An eleven-year review via longitudinal mixed effects modeling.

dc.type

Journal article

pubs.author-url

https://www.ncbi.nlm.nih.gov/pubmed/28480599

pubs.organisational-group

Clinical Science Departments

pubs.organisational-group

Duke

pubs.organisational-group

Duke Human Vaccine Institute

pubs.organisational-group

Institutes and Centers

pubs.organisational-group

Medicine

pubs.organisational-group

Medicine, Duke Human Vaccine Institute

pubs.organisational-group

School of Medicine

pubs.organisational-group

Staff

pubs.publication-status

Published online

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
PMID 28480599.pdf
Size:
340.14 KB
Format:
Adobe Portable Document Format