Implementing a Continuous Quality Improvement Program in a High-Volume Clinical Echocardiography Laboratory: Improving Care for Patients With Aortic Stenosis.

dc.contributor.author

Samad, Zainab

dc.contributor.author

Minter, Stephanie

dc.contributor.author

Armour, Alicia

dc.contributor.author

Tinnemore, Amanda

dc.contributor.author

Sivak, Joseph A

dc.contributor.author

Sedberry, Brenda

dc.contributor.author

Strub, Karen

dc.contributor.author

Horan, Seanna M

dc.contributor.author

Harrison, J Kevin

dc.contributor.author

Kisslo, Joseph

dc.contributor.author

Douglas, Pamela S

dc.contributor.author

Velazquez, Eric J

dc.coverage.spatial

United States

dc.date.accessioned

2017-10-02T20:02:13Z

dc.date.available

2017-10-02T20:02:13Z

dc.date.issued

2016-03

dc.description.abstract

BACKGROUND: The management of aortic stenosis rests on accurate echocardiographic diagnosis. Hence, it was chosen as a test case to examine the utility of continuous quality improvement (CQI) approaches to increase echocardiographic data accuracy and reliability. A novel, multistep CQI program was designed and prospectively used to investigate whether it could minimize the difference in aortic valve mean gradients reported by echocardiography when compared with cardiac catheterization. METHODS AND RESULTS: The Duke Echo Laboratory compiled a multidisciplinary CQI team including 4 senior sonographers and MD faculty to develop a mapped CQI process that incorporated Intersocietal Accreditation Commission standards. Quarterly, the CQI team reviewed all moderate- or greater-severity aortic stenosis echocardiography studies with concomitant catheterization data, and deidentified individual and group results were shared at meetings attended by cardiologists and sonographers. After review of 2011 data, the CQI team proposed specific amendments implemented over 2012: the use of nontraditional imaging and Doppler windows as well as evaluation of aortic gradients by a second sonographer. The primary outcome measure was agreement between catheterization- and echocardiography-derived mean gradients calculated by using the coverage probability index with a prespecified acceptable echocardiography-catheterization difference of <10 mm Hg in mean gradient. Between January 2011 and January 2014, 2093 echocardiograms reported moderate or greater aortic stenosis. Among cases with available catheterization data pre- and post-CQI, the coverage probability index increased from 54% to 70% (P=0.03; 98 cases, year 2011; 70 cases, year 2013). The proportion of patients referred for invasive valve hemodynamics decreased from 47% pre-CQI to 19% post-CQI (P<0.001). CONCLUSIONS: A laboratory practice pattern that was amenable to reform was identified, and a multistep modification was designed and implemented that produced clinically valuable performance improvements. The new protocol improved aortic stenosis mean gradient agreement between echocardiography and catheterization and was associated with a measurable decrease in referrals of patients for invasive studies.

dc.identifier

https://www.ncbi.nlm.nih.gov/pubmed/26957220

dc.identifier

CIRCIMAGING.115.003708

dc.identifier.eissn

1942-0080

dc.identifier.uri

https://hdl.handle.net/10161/15611

dc.language

eng

dc.publisher

Ovid Technologies (Wolters Kluwer Health)

dc.relation.ispartof

Circ Cardiovasc Imaging

dc.relation.isversionof

10.1161/CIRCIMAGING.115.003708

dc.subject

aortic valve stenosis

dc.subject

catheterization

dc.subject

echocardiography

dc.subject

image acquisition

dc.subject

image quality

dc.subject

quality improvement

dc.subject

Aortic Valve

dc.subject

Aortic Valve Stenosis

dc.subject

Cardiac Catheterization

dc.subject

Critical Pathways

dc.subject

Echocardiography, Doppler

dc.subject

Feasibility Studies

dc.subject

Hemodynamics

dc.subject

Hospitals, High-Volume

dc.subject

Humans

dc.subject

Observer Variation

dc.subject

Patient Care Team

dc.subject

Practice Patterns, Physicians'

dc.subject

Predictive Value of Tests

dc.subject

Program Evaluation

dc.subject

Prospective Studies

dc.subject

Quality Improvement

dc.subject

Quality Indicators, Health Care

dc.subject

Reproducibility of Results

dc.subject

Severity of Illness Index

dc.subject

Workflow

dc.title

Implementing a Continuous Quality Improvement Program in a High-Volume Clinical Echocardiography Laboratory: Improving Care for Patients With Aortic Stenosis.

dc.type

Journal article

duke.contributor.orcid

Douglas, Pamela S|0000-0001-9876-4049

duke.contributor.orcid

Velazquez, Eric J|0000-0003-2245-7477

pubs.author-url

https://www.ncbi.nlm.nih.gov/pubmed/26957220

pubs.issue

3

pubs.organisational-group

Clinical Science Departments

pubs.organisational-group

Duke

pubs.organisational-group

Duke Clinical Research Institute

pubs.organisational-group

Institutes and Centers

pubs.organisational-group

Medicine

pubs.organisational-group

Medicine, Cardiology

pubs.organisational-group

School of Medicine

pubs.organisational-group

Staff

pubs.publication-status

Published

pubs.volume

9

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
SH Publication.pdf
Size:
1.69 MB
Format:
Adobe Portable Document Format