Modified INOvent for delivery of inhaled nitric oxide during cardiac MRI.

Abstract

BACKGROUND: The aim of this study was to assess the feasibility of delivering NO through a modified system to allow clearance of the magnetic field and thus compatibility with cardiac magnetic resonance (CMR). Nitric oxide (NO) is an inhalational, selective pulmonary vasodilator with a wide range of applications in a variety of disease states, including diseases that affect the right ventricle. Accurate assessment of dynamic changes in right ventricular function necessitates CMR; however, delivery of NO is only possible using equipment that is not magnetic resonance imaging (MRI) compatible (INOvent delivery system, Ohmeda, Inc., Madison, WI, USA). METHODS: The INOvent delivery system was modified by using 35 ft. of standard oxygen tubing to allow NO delivery through an electrical conduit and into the MRI suite. The concentrations of oxygen (O(2)), nitrogen dioxide (a harmful byproduct, NO(2)) and NO were measured in triplicate using the built-in electrochemical analyzer on the INOvent. After confirmation of safety, the system was used to administer drug to a patient x, and dynamic MRI measurements were performed. RESULTS: When the standard INOvent was set to administer 40 ppm of NO, the mean/standard deviation of gas delivered was as follows: NO: 42/0 ppm; NO(2): 0.3/0.1 ppm; and O(2): 93/0 ppm. In comparison, the gas delivery of the modified INOvent was follows: NO: 41/0 ppm; NO(2): 0.5/0 ppm; and O(2): 93.7/0.6 ppm. During administration to an index patient with severe pulmonic insufficiency (PI), a measurable reduction in PI was observed by CMR. CONCLUSIONS: Nitric oxide can be administered through 35 ft. of standard oxygen tubing without significantly affecting dose delivery. This technique has potential application in patients with right-sided structural heart disease for determination of dynamic physiological changes.

Department

Description

Provenance

Citation

Published Version (Please cite this version)

10.1016/j.mri.2011.05.008

Publication Info

Devendra, Ganesh P, Stephen A Hart, Yuli Y Kim, Randy M Setser, Scott D Flamm and Richard A Krasuski (2011). Modified INOvent for delivery of inhaled nitric oxide during cardiac MRI. Magn Reson Imaging, 29(8). pp. 1145–1149. 10.1016/j.mri.2011.05.008 Retrieved from https://hdl.handle.net/10161/11025.

This is constructed from limited available data and may be imprecise. To cite this article, please review & use the official citation provided by the journal.


Unless otherwise indicated, scholarly articles published by Duke faculty members are made available here with a CC-BY-NC (Creative Commons Attribution Non-Commercial) license, as enabled by the Duke Open Access Policy. If you wish to use the materials in ways not already permitted under CC-BY-NC, please consult the copyright owner. Other materials are made available here through the author’s grant of a non-exclusive license to make their work openly accessible.