Background

Proximal junctional kyphosis (PJK) is a common postoperative complication after adult spinal deformity (ASD) surgery and may manifest with neurological decline, worsening spinal deformity, and spinal instability, which warrant reoperation. Rates of PJK may be as high as 69.4% after ASD surgery.

Objective

To evaluate the efficacy of junctional tethers for PJK prophylaxis after multilevel instrumented surgery for ASD with minimum 2-yr follow-up.

Methods

Single-center retrospective analysis of adult patients (age ≥18 yr) who underwent ASD surgery with index operations performed between November 2010 and June 2016 and achieved minimum 2-yr follow-up. Patients with ASD were subdivided into 3 treatment cohorts based on institutional protocol: no tether (NT), polyethylene tether-only (TO), and tether with crosslink (TC). PJK was defined as a proximal junctional angle (PJA) >10° and 10° greater than the corresponding preoperative measurement. Patient demographics, operative details, standard radiographic scoliosis measurements (including PJA and assessment of PJK), and complications were analyzed.

Results

Of 184 patients, 146 (79.3%) achieved minimum 2-yr follow-up (mean = 45 mo; mean age = 67 yr; 67.8% women). PJK rates reported for the NT, TO, and TC cohorts were 60.7% (37/61), 35.7% (15/42), and 23.3% (10/43), respectively. PJK rates among TC patients were significantly lower than NT (P = .01601).

Conclusion

Junctional tethers with crosslink significantly reduced the incidence of PJK and revisions for PJK among ASD patients treated with long-segment posterior instrumented fusions who achieved minimum 2-yr follow-up.

Department

Description

Provenance

Citation

Published Version (Please cite this version)

10.1093/neuros/nyab123

Publication Info

Rabinovich, Emily P, M Harrison Snyder, Jesse J McClure, Thomas J Buell, Justin S Smith, Christopher I Shaffrey and Avery L Buchholz (2021). Posterior Polyethylene Tethers Reduce Occurrence of Proximal Junctional Kyphosis After Multilevel Spinal Instrumentation for Adult Spinal Deformity: A Retrospective Analysis. Neurosurgery, 89(2). pp. 227–235. 10.1093/neuros/nyab123 Retrieved from https://hdl.handle.net/10161/28085.

This is constructed from limited available data and may be imprecise. To cite this article, please review & use the official citation provided by the journal.

Scholars@Duke

Shaffrey

Christopher Ignatius Shaffrey

Professor of Orthopaedic Surgery

I have more than 25 years of experience treating patients of all ages with spinal disorders. I have had an interest in the management of spinal disorders since starting my medical education. I performed residencies in both orthopaedic surgery and neurosurgery to gain a comprehensive understanding of the entire range of spinal disorders. My goal has been to find innovative ways to manage the range of spinal conditions, straightforward to complex. I have a focus on managing patients with complex spinal disorders. My patient evaluation and management philosophy is to provide engaged, compassionate care that focuses on providing the simplest and least aggressive treatment option for a particular condition. In many cases, non-operative treatment options exist to improve a patient’s symptoms. I have been actively engaged in clinical research to find the best ways to manage spinal disorders in order to achieve better results with fewer complications.


Unless otherwise indicated, scholarly articles published by Duke faculty members are made available here with a CC-BY-NC (Creative Commons Attribution Non-Commercial) license, as enabled by the Duke Open Access Policy. If you wish to use the materials in ways not already permitted under CC-BY-NC, please consult the copyright owner. Other materials are made available here through the author’s grant of a non-exclusive license to make their work openly accessible.