Goal-Driven Beam Setting Optimization for Whole-Breast Radiation Therapy.

dc.contributor.author

Wang, Wentao

dc.contributor.author

Sheng, Yang

dc.contributor.author

Yoo, Sua

dc.contributor.author

Blitzblau, Rachel C

dc.contributor.author

Yin, Fang-Fang

dc.contributor.author

Wu, Q Jackie

dc.date.accessioned

2019-10-01T14:03:38Z

dc.date.available

2019-10-01T14:03:38Z

dc.date.issued

2019-01

dc.date.updated

2019-10-01T14:03:37Z

dc.description.abstract

PURPOSE:To develop an automated optimization program to generate optimal beam settings for whole-breast radiation therapy driven by clinically oriented goals. MATERIALS AND METHODS:Forty patients were retrospectively included in this study. Each patient's planning images, contoured structures of planning target volumes, organs-at-risk, and breast wires were used to optimize for patient-specific-beam settings. Two beam geometries were available tangential beams only and tangential plus supraclavicular beams. Beam parameters included isocenter position, gantry, collimator, couch angles, and multileaf collimator shape. A geometry-based goal function was defined to determine such beam parameters to minimize out-of-field target volume and in-field ipsilateral lung volume. For each geometry, the weighting in the goal function was trained with 10 plans and tested on 10 additional plans. For each query patient, the optimal beam setting was searched for different gantry-isocenter pairs. Optimal fluence maps were generated by an in-house automatic fluence optimization program for target coverage and homogeneous dose distribution, and dose calculation was performed in Eclipse. Automatically generated plans were compared with manually generated plans for target coverage and lung and heart sparing. RESULTS:The program successfully produced a set of beam parameters for every patient. Beam optimization time ranged from 10 to 120 s. The automatic plans had overall comparable plan quality to manually generated plans. For all testing cases, the mean target V95% was 91.0% for the automatic plans and 88.5% for manually generated plans. The mean ipsilateral lung V20Gy was lower for the automatic plans (15.2% vs 17.9%). The heart mean dose, maximum dose of the body, and conformity index were all comparable. CONCLUSION:We developed an automated goal-driven beam setting optimization program for whole-breast radiation therapy. It provides clinically relevant solutions based on previous clinical practice as well as patient specific anatomy on a substantially faster time frame.

dc.identifier.issn

1533-0346

dc.identifier.issn

1533-0338

dc.identifier.uri

https://hdl.handle.net/10161/19367

dc.language

eng

dc.publisher

SAGE Publications

dc.relation.ispartof

Technology in cancer research & treatment

dc.relation.isversionof

10.1177/1533033819858661

dc.subject

automation

dc.subject

beam geometry

dc.subject

breast cancer

dc.subject

optimization

dc.subject

treatment planning

dc.subject

whole breast radiation therapy

dc.title

Goal-Driven Beam Setting Optimization for Whole-Breast Radiation Therapy.

dc.type

Journal article

duke.contributor.orcid

Sheng, Yang|0000-0003-3380-1966

duke.contributor.orcid

Blitzblau, Rachel C|0000-0002-4296-2238

duke.contributor.orcid

Yin, Fang-Fang|0000-0002-2025-4740|0000-0003-1064-2149

pubs.begin-page

1533033819858661

pubs.organisational-group

School of Medicine

pubs.organisational-group

Duke

pubs.organisational-group

Radiation Oncology

pubs.organisational-group

Clinical Science Departments

pubs.organisational-group

Duke Cancer Institute

pubs.organisational-group

Institutes and Centers

pubs.organisational-group

Duke Kunshan University Faculty

pubs.organisational-group

Duke Kunshan University

pubs.publication-status

Published

pubs.volume

18

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Goal-Driven Beam Setting Optimization for Whole-Breast Radiation Therapy.pdf
Size:
791.5 KB
Format:
Adobe Portable Document Format