PURPOSE:MRI-based treatment planning is a promising technique for liver stereotactic-body radiation therapy (SBRT) treatment planning to improve target volume delineation and reduce radiation dose to normal tissues. MR geometric distortion, however, is a source of potential error in MRI-based treatment planning. The aim of this study is to investigate dosimetric uncertainties caused by MRI geometric distortion in MRI-based treatment planning for liver SBRT. MATERIALS AND METHODS:The study was conducted using computer simulations. 3D MR geometric distortion was simulated using measured data in the literature. Planning MR images with distortions were generated by integrating the simulated 3D MR geometric distortion onto planning CT images. MRI-based treatment plans were then generated on the planning MR images with two dose calculation methods: (1) using original CT numbers; and (2) using organ-specific assigned CT numbers. Dosimetric uncertainties of various dose-volume-histogram parameters were determined as their differences between the simulated MRI-based plans and the original clinical CT-based plans for five liver SBRT cases. RESULTS:The average simulated distortion for the five liver SBRT cases was 2.77 mm. In the case of using original CT numbers for dose calculation, the average dose uncertainties for target volumes and critical structures were <0.5 Gy, and the average target volume percentage at prescription dose uncertainties was 0.97%. In the case of using assigned CT numbers, the average dose uncertainties for target volumes and critical structures were <1.0 Gy, and the average target volume percentage at prescription dose uncertainties was 2.02%. CONCLUSIONS:Dosimetric uncertainties caused by MR geometric distortion in MRI-based liver SBRT treatment planning was generally small (<1 Gy) when the distortion is 3 mm.

dc.identifier.issn

1526-9914

dc.identifier.issn

1526-9914

dc.identifier.uri

https://hdl.handle.net/10161/19371

dc.language

eng

dc.publisher

Wiley

dc.relation.ispartof

Journal of applied clinical medical physics

dc.relation.isversionof

10.1002/acm2.12520

dc.subject

Humans

dc.subject

Liver Neoplasms

dc.subject

Magnetic Resonance Imaging

dc.subject

Radiosurgery

dc.subject

Radiotherapy Dosage

dc.subject

Radiotherapy Planning, Computer-Assisted

dc.subject

Uncertainty

dc.subject

Phantoms, Imaging

dc.subject

Image Processing, Computer-Assisted

dc.subject

Radiotherapy, Intensity-Modulated

dc.subject

Organs at Risk

dc.title

Evaluation of dosimetric uncertainty caused by MR geometric distortion in MRI-based liver SBRT treatment planning.

dc.type

Journal article

duke.contributor.orcid

Yin, Fang-Fang|0000-0002-2025-4740|0000-0003-1064-2149

pubs.begin-page

43

pubs.end-page

50

pubs.issue

2

pubs.organisational-group

School of Medicine

pubs.organisational-group

Duke

pubs.organisational-group

Radiation Oncology

pubs.organisational-group

Clinical Science Departments

pubs.organisational-group

Duke Kunshan University Faculty

pubs.organisational-group

Duke Kunshan University

pubs.organisational-group

Duke Cancer Institute

pubs.organisational-group

Institutes and Centers

pubs.publication-status

Published

pubs.volume

20

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Evaluation of dosimetric uncertainty caused by MR geometric distortion in MRI-based liver SBRT treatment planning.pdf
Size:
637.36 KB
Format:
Adobe Portable Document Format