Automated quality control in nuclear medicine using the structured noise index.
Date
2020-04
Authors
Journal Title
Journal ISSN
Volume Title
Repository Usage Stats
views
downloads
Citation Stats
Attention Stats
Abstract
Purpose
Daily flood-field uniformity evaluation serves as the central element of nuclear medicine (NM) quality control (QC) programs. Uniformity images are traditionally analyzed using pixel value-based metrics, that is, integral uniformity (IU), which often fail to capture subtle structure and patterns caused by changes in gamma camera performance, requiring visual inspections which are subjective and time demanding. The goal of this project was to implement an advanced QC metrology for NM to effectively identify nonuniformity issues, and report issues in a timely manner for efficient correction prior to clinical use. The project involved the implementation of the program over a 2-year period at a multisite major medical institution.Methods
Using a previously developed quantitative uniformity analysis metric, the structured noise index (SNI) [Nelson et al. (2014), \textit{J Nucl Med.}, \textbf{55}:169-174], an automated QC process was developed to analyze, archive, and report on daily NM QC uniformity images. Clinical implementation of the newly developed program ran in parallel with the manufacturer's reported IU-based QC program. The effectiveness of the SNI program was evaluated over a 21-month period using sensitivity and coefficient of variation statistics.Results
A total of 7365 uniformity QC images were analyzed. Lower level SNI alerts were generated in 12.5% of images and upper level alerts in 1.7%. Intervention due to image quality issues occurred on 26 instances; the SNI metric identified 24, while the IU metric identified eight. The SNI metric reported five upper level alerts where no clinical engineering intervention was deemed necessary.Conclusion
An SNI-based QC program provides a robust quantification of the performance of gamma camera uniformity. It operates seamlessly across a fleet of multiple camera models and, additionally, provides effective workflow among the clinical staff. The reliability of this process could eliminate the need for visual inspection of each image, saving valuable time, while enabling quantitative analysis of inter- and intrasystem performance.Type
Department
Description
Provenance
Subjects
Citation
Permalink
Published Version (Please cite this version)
Publication Info
Nelson, Jeffrey S, and Ehsan Samei (2020). Automated quality control in nuclear medicine using the structured noise index. Journal of applied clinical medical physics, 21(4). pp. 80–86. 10.1002/acm2.12850 Retrieved from https://hdl.handle.net/10161/26095.
This is constructed from limited available data and may be imprecise. To cite this article, please review & use the official citation provided by the journal.
Collections
Scholars@Duke

Jeffrey Nelson
Unless otherwise indicated, scholarly articles published by Duke faculty members are made available here with a CC-BY-NC (Creative Commons Attribution Non-Commercial) license, as enabled by the Duke Open Access Policy. If you wish to use the materials in ways not already permitted under CC-BY-NC, please consult the copyright owner. Other materials are made available here through the author’s grant of a non-exclusive license to make their work openly accessible.