Influences of interpolation error, electrode geometry, and the electrode-tissue interface on models of electric fields produced by deep brain stimulation.

dc.contributor.author

Howell, Bryan

dc.contributor.author

Naik, Sagar

dc.contributor.author

Grill, Warren M

dc.date.accessioned

2021-09-28T18:54:53Z

dc.date.available

2021-09-28T18:54:53Z

dc.date.issued

2014-02

dc.date.updated

2021-09-28T18:54:52Z

dc.description.abstract

Deep brain stimulation (DBS) is an established therapy for movement disorders, but the fundamental mechanisms by which DBS has its effects remain unknown. Computational models can provide insights into the mechanisms of DBS, but to be useful, the models must have sufficient detail to predict accurately the electric fields produced by DBS. We used a finite-element method model of the Medtronic 3387 electrode array, coupled to cable models of myelinated axons, to quantify how interpolation errors, electrode geometry, and the electrode-tissue interface affect calculation of electrical potentials and stimulation thresholds for populations of model nerve fibers. Convergence of the potentials was not a sufficient criterion for ensuring the same degree of accuracy in subsequent determination of stimulation thresholds, because the accuracy of the stimulation thresholds depended on the order of the elements. Simplifying the 3387 electrode array by ignoring the inactive contacts and extending the terminated end of the shaft had position-dependent effects on the potentials and excitation thresholds, and these simplifications may impact correlations between DBS parameters and clinical outcomes. When the current density in the bulk tissue is uniform, the effect of the electrode-tissue interface impedance could be approximated by filtering the potentials calculated with a static lumped electrical equivalent circuit. Further, for typical DBS parameters during voltage-regulated stimulation, it was valid to approximate the electrode as an ideal polarized electrode with a nonlinear capacitance. Validation of these computational considerations enables accurate modeling of the electric field produced by DBS.

dc.identifier.issn

0018-9294

dc.identifier.issn

1558-2531

dc.identifier.uri

https://hdl.handle.net/10161/23857

dc.language

eng

dc.publisher

Institute of Electrical and Electronics Engineers (IEEE)

dc.relation.ispartof

IEEE transactions on bio-medical engineering

dc.relation.isversionof

10.1109/tbme.2013.2292025

dc.subject

Humans

dc.subject

Deep Brain Stimulation

dc.subject

Electrodes

dc.subject

Biomedical Engineering

dc.subject

Finite Element Analysis

dc.subject

Models, Neurological

dc.title

Influences of interpolation error, electrode geometry, and the electrode-tissue interface on models of electric fields produced by deep brain stimulation.

dc.type

Journal article

duke.contributor.orcid

Howell, Bryan|0000-0002-3329-8478

pubs.begin-page

297

pubs.end-page

307

pubs.issue

2

pubs.organisational-group

Pratt School of Engineering

pubs.organisational-group

Biomedical Engineering

pubs.organisational-group

Electrical and Computer Engineering

pubs.organisational-group

Neurobiology

pubs.organisational-group

Duke Science & Society

pubs.organisational-group

Duke Institute for Brain Sciences

pubs.organisational-group

Neurosurgery

pubs.organisational-group

Duke

pubs.organisational-group

Basic Science Departments

pubs.organisational-group

School of Medicine

pubs.organisational-group

Initiatives

pubs.organisational-group

Institutes and Provost's Academic Units

pubs.organisational-group

University Institutes and Centers

pubs.organisational-group

Clinical Science Departments

pubs.publication-status

Published

pubs.volume

61

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
nihms567410.pdf
Size:
2.11 MB
Format:
Adobe Portable Document Format
Description:
Accepted version