A systematic assessment and optimization of photon-counting CT for lung density quantifications.

dc.contributor.author

Sotoudeh-Paima, Saman

dc.contributor.author

Segars, W Paul

dc.contributor.author

Ghosh, Dhrubajyoti

dc.contributor.author

Luo, Sheng

dc.contributor.author

Samei, Ehsan

dc.contributor.author

Abadi, Ehsan

dc.date.accessioned

2024-03-01T15:27:14Z

dc.date.available

2024-03-01T15:27:14Z

dc.date.issued

2024-02

dc.description.abstract

Background

Photon-counting computed tomography (PCCT) has recently emerged into clinical use; however, its optimum imaging protocols and added benefits remains unknown in terms of providing more accurate lung density quantification compared to energy-integrating computed tomography (EICT) scanners.

Purpose

To systematically assess the performance of a clinical PCCT scanner for lung density quantifications and compare it against EICT.

Methods

This cross-sectional study involved a retrospective analysis of subjects scanned (August-December 2021) using a clinical PCCT system. The influence of altering reconstruction parameters was studied (reconstruction kernel, pixel size, slice thickness). A virtual CT dataset of anthropomorphic virtual subjects was acquired to demonstrate the correspondence of findings to clinical dataset, and to perform systematic imaging experiments, not possible using human subjects. The virtual subjects were imaged using a validated, scanner-specific CT simulator of a PCCT and two EICT (defined as EICT A and B) scanners. The images were evaluated using mean absolute error (MAE) of lung and emphysema density against their corresponding ground truth.

Results

Clinical and virtual PCCT datasets showed similar trends, with sharper kernels and smaller voxel sizes increasing percentage of low-attenuation areas below -950 HU (LAA-950) by up to 15.7 ± 6.9% and 11.8 ± 5.5%, respectively. Under the conditions studied, higher doses, thinner slices, smaller pixel sizes, iterative reconstructions, and quantitative kernels with medium sharpness resulted in lower lung MAE values. While using these settings for PCCT, changes in the dose level (13 to 1.3 mGy), slice thickness (0.4 to 1.5 mm), pixel size (0.49 to 0.98 mm), reconstruction technique (70 keV-VMI to wFBP), and kernel (Qr48 to Qr60) increased lung MAE by 15.3 ± 2.0, 1.4 ± 0.6, 2.2 ± 0.3, 4.2 ± 0.8, and 9.1 ± 1.6 HU, respectively. At the optimum settings identified per scanner, PCCT images exhibited lower lung and emphysema MAE than those of EICT scanners (by 2.6 ± 1.0 and 9.6 ± 3.4 HU, compared to EICT A, and by 4.8 ± 0.8 and 7.4 ± 2.3 HU, compared to EICT B). The accuracy of lung density measurements was correlated with subjects' mean lung density (p < 0.05), measured by PCCT at optimum setting under the conditions studied.

Conclusion

Photon-counting CT demonstrated superior performance in density quantifications, with its influences of imaging parameters in line with energy-integrating CT scanners. The technology offers improvement in lung quantifications, thus demonstrating potential toward more objective assessment of respiratory conditions.
dc.identifier.issn

0094-2405

dc.identifier.issn

2473-4209

dc.identifier.uri

https://hdl.handle.net/10161/30207

dc.language

eng

dc.publisher

Wiley

dc.relation.ispartof

Medical physics

dc.relation.isversionof

10.1002/mp.16987

dc.rights.uri

https://creativecommons.org/licenses/by-nc/4.0

dc.subject

computed tomography

dc.subject

photon-counting CT

dc.subject

quantitative imaging

dc.subject

virtual imaging trials

dc.title

A systematic assessment and optimization of photon-counting CT for lung density quantifications.

dc.type

Journal article

duke.contributor.orcid

Luo, Sheng|0000-0003-4214-5809

duke.contributor.orcid

Abadi, Ehsan|0000-0002-9123-5854

pubs.organisational-group

Duke

pubs.organisational-group

Pratt School of Engineering

pubs.organisational-group

School of Medicine

pubs.organisational-group

Basic Science Departments

pubs.organisational-group

Clinical Science Departments

pubs.organisational-group

Institutes and Centers

pubs.organisational-group

Biostatistics & Bioinformatics

pubs.organisational-group

Electrical and Computer Engineering

pubs.organisational-group

Radiology

pubs.organisational-group

Duke Clinical Research Institute

pubs.organisational-group

Biostatistics & Bioinformatics, Division of Biostatistics

pubs.publication-status

Published

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
2024Sotoudeh‐Paima_et_al2024Medical Physics.pdf
Size:
4.8 MB
Format:
Adobe Portable Document Format
Description:
Published version