Evaluation of an electron Monte Carlo dose calculation algorithm for electron beam.
Date
2008-06-23
Journal Title
Journal ISSN
Volume Title
Repository Usage Stats
views
downloads
Citation Stats
Abstract
The electron Monte Carlo (eMC) dose calculation algorithm of the Eclipse treatment planning system is based heavily upon Monte Carlo simulation of the linac head and modeling of the linac beam characteristics with minimal measurement of beam data. Commissioning of the eMC algorithm on multiple identical linacs provided a unique opportunity to systematically evaluate the algorithm with actual measurements of clinically relevant beam and dose parameters. In this study, measured and eMC calculated dose distributions were compared both along and perpendicular to electron beam direction for electron energy/applicator/depth combination using measurement data from four Varian 21EX CLINAC linear accelerator (Varian Medical System, Palo Alto, CA). Cutout factors for sizes down to 3 x 3 cm were also compared. Comparisons between the measurement and the eMC calculated values show that the R90, R80, R50, and R10 values mostly agree within 3 mm. Measure and Calculated bremsstrahlung dose Dx correlates well statistically although eMC calculated Dx values are consistently smaller than the measured, with maximum discrepancy of 1% for the 20 MeV electron beams. Surface dose agrees mostly within 2%. Field width and penumbra agree mostly within 3mm. Calculation grid size is found to have a significant effect on the dose calculation. A grid size of 5 mm can produce erroneous dose distributions. Using a grid size of 2.5 mm and a 3% accuracy specified for the eMC to stop calculation iteration, the absolute output agrees with measurements within 3% for field sizes of 5 x 5 cm or larger. For cutout of 3 x 3 cm, however, the output disagreement can reach 8%. Our result indicate that eMC algorithm in Eclipse provides acceptable agreement with measurement data for most clinical situations. Calculation grid size of 2.5 mm or smaller is recommended.
Type
Department
Description
Provenance
Citation
Permalink
Published Version (Please cite this version)
Publication Info
Hu, Ye Angela, Haijun Song, Zhe Chen, Sumin Zhou and Fang-Fang Yin (2008). Evaluation of an electron Monte Carlo dose calculation algorithm for electron beam. Journal of applied clinical medical physics, 9(3). p. 2720. 10.1120/jacmp.v9i3.2720 Retrieved from https://hdl.handle.net/10161/19393.
This is constructed from limited available data and may be imprecise. To cite this article, please review & use the official citation provided by the journal.
Collections
Scholars@Duke
Fang-Fang Yin
Stereotactic radiosurgery, Stereotactic body radiation therapy, treatment planning optimization, knowledge guided radiation therapy, intensity-modulated radiation therapy, image-guided radiation therapy, oncological imaging and informatics
Unless otherwise indicated, scholarly articles published by Duke faculty members are made available here with a CC-BY-NC (Creative Commons Attribution Non-Commercial) license, as enabled by the Duke Open Access Policy. If you wish to use the materials in ways not already permitted under CC-BY-NC, please consult the copyright owner. Other materials are made available here through the author’s grant of a non-exclusive license to make their work openly accessible.